Analysis of Retrieved Hydroxyapatite-Coated Hip Prostheses

作者: K.A. Gross , W. Walsh , E. Swarts

DOI: 10.1361/10599630418112

关键词:

摘要: Hydroxyapatite (HAp) coatings are used extensively on orthopaedic devices to improve the adhesion of bone prosthesis. This approach increases integrity and longevity implanted Four HAp-coated hip components recovered from patients during revision surgery were investigated for attachment coating modification after storage in ethanol or formaldehyde. Orthopedic displayed preferable microtextured areas little smoother areas. The microstructure differed between three that remained prosthesis surface, ranging completely crystalline made by vacuum plasma spraying less manufactured air spraying. Coating failure lower crystallinity was observed a crack at interface possibly caused dissolution an amorphous phase. While higher degraded delamination, produced loose particulate outer surface. morphology as scanning electron microscopy (SEM) lamellae fracture, chemical dissolution, osteoclastic resorption, precipitation agreement with previously identified in-vitro events. appeared be extended those subject levels stress more coverage.

参考文章(78)
Harry B. Skinner, Agnes S. Kim, Joyce H. Keyak, C. D. Mote, Femoral prosthesis implantation induces changes in bone stress that depend on the extent of porous coating. Journal of Orthopaedic Research. ,vol. 12, pp. 553- 563 ,(1994) , 10.1002/JOR.1100120412
Ioana Baltag, Kouichi Watanabe, Haruka Kusakari, Naoyuki Taguchi, Osamu Miyakawa, Masayoshi Kobayashi, Naoko Ito, Long-term changes of hydroxyapatite-coated dental implants. Journal of Biomedical Materials Research. ,vol. 53, pp. 76- 85 ,(2000) , 10.1002/(SICI)1097-4636(2000)53:1<76::AID-JBM11>3.0.CO;2-4
Michael A. Morris, Patrick J. Kelly, Use of tracer microspheres to measure bone blood flow in conscious dogs Calcified Tissue International. ,vol. 32, pp. 69- 76 ,(1980) , 10.1007/BF02408523
Ravishankar Vedantam, C. Ruddlesdin, The fully hydroxyapatite-coated total hip implant. Clinical and roentgenographic results Journal of Arthroplasty. ,vol. 11, pp. 534- 542 ,(1996) , 10.1016/S0883-5403(96)80106-9
CHARLES A. ENGH, J. DENNIS BOBYN, The influence of stem size and extent of porous coating on femoral bone resorption after primary cementless hip arthroplasty. Clinical Orthopaedics and Related Research. ,vol. 231, pp. 7- 28 ,(1988) , 10.1097/00003086-198806000-00002
John P. Collier, Victor A. Surprenant, Michael B. Mayor, Marguerite Wrona, Robert E. Jensen, Helene P. Surprenant, Loss of hydroxyapatite coating on retrieved, total hip components. Journal of Arthroplasty. ,vol. 8, pp. 389- 393 ,(1993) , 10.1016/S0883-5403(06)80037-9
K. A. Gross, C. C. Berndt, Biomedical Application of Apatites Reviews in Mineralogy & Geochemistry. ,vol. 48, pp. 631- 672 ,(2002) , 10.2138/RMG.2002.48.17
J. D. de Bruijn, Y. P. Bovell, J. E. Davies, C. A. van Blitterswijk, Osteoclastic resorption of calcium phosphates is potentiated in postosteogenic culture conditions Journal of Biomedical Materials Research. ,vol. 28, pp. 105- 112 ,(1994) , 10.1002/JBM.820280114
Pieter Buma, Jean W.M. Gardeniers, Tissue reactions around a hydroxyapatite-coated hip prosthesis. Case report of a retrieved specimen. Journal of Arthroplasty. ,vol. 10, pp. 389- 395 ,(1995) , 10.1016/S0883-5403(05)80190-1