Closed loop control of sedation for colonoscopy using the Bispectral Index*

作者: K. Leslie , A. Absalom , G. N. C. Kenny

DOI: 10.1046/J.1365-2044.2002.02572_2.X

关键词:

摘要: Summary Sixteen patients undergoing colonoscopy were sedated with propofol using a closed-loop control system guided by the Bispectral Index (BIS). Propofol administration, via target-controlled infusion, was controlled proportional-integral-differential algorithm. The median (range) target concentration during 2.3 (1.7–3.6) µg.ml−1. performance characteristics of excellent, absolute error 7 (1–15)%. Patients drowsy yet rousable, BIS set-point 80 (75–85). No patient became apnoeic, required airway support or haemodynamically unstable whilst sedated. Eight moaned moved and four had recall. Median time to full consciousness 4 (2–20) min after end control. Patient surgeon satisfaction high. We conclude that may be suitable variable for sedation propofol.

参考文章(19)
DORIS A. CHERNIK, DENNIS GILLINGS, HARRIET LAINE, JUDITH HENDLER, JONATHAN M. SILVER, ARNOLD B. DAVIDSON, ELIAS M. SCHWAM, JUDITH L. SIEGEL, Validity and reliability of the Observer's Assessment of Alertness/Sedation Scale: study with intravenous midazolam. Journal of Clinical Psychopharmacology. ,vol. 10, pp. 244- 251 ,(1990) , 10.1097/00004714-199008000-00003
Julian F. Biebuyck, Dorene A. O'Hara, Daniel K. Bogen, Abraham Noordergraaf, The use of computers for controlling the delivery of anesthesia. Anesthesiology. ,vol. 77, pp. 563- 581 ,(1992) , 10.1097/00000542-199209000-00024
J. ANTONIO ALDRETE, DIANE KROULIK, A postanesthetic recovery score. Anesthesia & Analgesia. ,vol. 49, pp. 924- 934 ,(1970) , 10.1213/00000539-197011000-00020
Y. Kawano, M. lmai, Y. Komura, N. Kikuchi, O. Kemmotsu, TRACHEAL CUFF AS A NEW CORE TEMPERATURE SITE Anesthesiology. ,vol. 77, pp. A563- ,(1992) , 10.1097/00000542-199209001-00563
Jay W. Johansen, Peter S. Sebel, Dennis M. Fisher, Development and Clinical Application of Electroencephalographic Bispectrum Monitoring Anesthesiology. ,vol. 93, pp. 1336- 1344 ,(2000) , 10.1097/00000542-200011000-00029
Peter S. A. Glass, Ira J. Rampil, Automated AnesthesiaFact or Fantasy Anesthesiology. ,vol. 95, pp. 1- 2 ,(2001) , 10.1097/00000542-200107000-00005
John R. Varvel, David L. Donoho, Steven L. Shafer, Measuring the predictive performance of computer-controlled infusion pumps Journal of Pharmacokinetics and Biopharmaceutics. ,vol. 20, pp. 63- 94 ,(1992) , 10.1007/BF01143186
Michel M. R. F. Struys, Tom De Smet, Linda F. M. Versichelen, Stijn Van de Velde, Rudy Van den Broecke, Eric P. Mortier, Comparison of Closed-loop Controlled Administration of Propofol Using Bispectral Index as the Controlled Variable versus “Standard Practice” Controlled Administration Anesthesiology. ,vol. 95, pp. 6- 17 ,(2001) , 10.1097/00000542-200107000-00007
E. Mortier, M. Struys, T. De Smet, L. Versichelen, G. Rolly, Closed‐loop controlled administration of propofol using bispectral analysis Anaesthesia. ,vol. 53, pp. 749- 754 ,(1998) , 10.1046/J.1365-2044.1998.00467.X
A. Morley, J. Derrick, P. Mainland, B. B. Lee, T. G. Short, Closed loop control of anaesthesia: an assessment of the bispectral index as the target of control Anaesthesia. ,vol. 55, pp. 953- 959 ,(2000) , 10.1046/J.1365-2044.2000.01527.X