Determination of chlorinated pesticide residues in drinking water by liquid-solid extraction and vacuum manifold system

作者: Gündüz Çiftçioğlu , Deniz Karlık , Metin Ateş

DOI: 10.5505/TURKHIJYEN.2015.52385

关键词:

摘要: Objective: The aim of this study is to determine the chlorinated pesticide residues in drinking water using a vacuum manifold system. Method: liquid- solid extraction method and GC/ µECD were used for determination pesticides water. Results: Chlorinated analysis are extracted from 1 L sample by passing through cartridge with C 18 organic phase. passed system at 4 mmHg pressure. This extract concentrated further evaporation. It was found that recovery values increased 53.7% 95.6%. precision 4.1% 19.8%. In calculating limit quantitation, standard deviation's numerical value which calculated each parameter experimentally multiplied 10. Analyzing the, Regulation on Concerning Water Indended Human Consumption permitted maximum residue limits Aldrin, Dieldrin, Heptaklor Heptaklorepoksid 0.03 µg/L it's 0.1 other parameters. reporting tested between 0.01 0.07 µg/L. Beta HCH (beta hexachlorocyclohexane) as

参考文章(4)
Henrik T. Rasmussen, Weiyong Li, Dirk Redlich, M. Ilias Jimidar, 6 – HPLC Method Development Separation Science and Technology. ,vol. 6, pp. 145- 190 ,(2005) , 10.1016/S0149-6395(05)80050-9
Marie-Claire Hennion, Solid-phase extraction: method development, sorbents, and coupling with liquid chromatography Journal of Chromatography A. ,vol. 856, pp. 3- 54 ,(1999) , 10.1016/S0021-9673(99)00832-8
Lloyd R. Snyder, Joseph L. Glajch, Joseph Jack Kirkland, Practical HPLC method development ,(1988)
Ronald E. Majors, Trends in Sample Preparation Lc Gc North America. ,vol. 10, pp. 1098- 1113 ,(1992)