Spin warp NMR imaging and applications to human whole-body imaging.

作者: G Johnson , T Redpath , W A Edelstein , J M S Hutchison

DOI: 10.1088/0031-9155/25/4/017

关键词:

摘要: Describes a new nuclear magnetic resonance (NMR) imaging technique which the authors call 'spin warp imaging' and gives examples of its application to human whole-body imaging. The apparatus is based on four-coil, air cored magnet (made by Oxford Instrument Company) capable accepting whole body. produces static field 0.04 T giving proton NMR frequency 1.7 MHz. maximum inhomogeneity about 6*10-4 at radius 0.23 m, approximately twice amount theoretically attainable with this configuration. pulse sequence used shown.

参考文章(2)
J M S Hutchinson, R J Sutherland, J R Mallard, NMR imaging: image recovery under magnetic fields with large non-uniformities. Journal of Physics E: Scientific Instruments. ,vol. 11, pp. 217- 222 ,(1978) , 10.1088/0022-3735/11/3/012
R. Damadian, Tumor Detection by Nuclear Magnetic Resonance Science. ,vol. 171, pp. 1151- 1153 ,(1971) , 10.1126/SCIENCE.171.3976.1151