DYSFUNCTION OF IMPLANTABLE DEFIBRILLATORS CAUSED BY SLOT MACHINES

作者: Antonio Madrid , Andres Sanchez , Enrique Bosch , Elíseo Fernández , Concepcion Moro Serrano

DOI: 10.1111/J.1540-8159.1997.TB04845.X

关键词:

摘要: The electrogram storage is a feature offered by the last generation of defibrillators, which enables physician to establish whether sensed episode and electrical therapy delivered device was appropriate. incidence electromagnetic interferences may be higher than initially suspected with latest defibrillators. Such interference documented in 4 out 100 patients over 24-month period. occurred all four them, while they were playing slot machines. In RR interval history, irregular QRS sensing variable frequency some episodes. intervals nonphysiological satisfying rate cut-off criteria an intermittent nonsustained way. There high risk between machines Our experience calls for warning defibrillator and, perhaps regulatory intervention.

参考文章(10)
Johannes Brachmann, Laurence D. Sterns, Thomas Hilbel, Thorsten Beyer, Wolfgang Schoels, Kirstin Freigang, Juraj Melichercik, Jutta Ruf-Richter, Wolfgang Kübler, Advances in follow-up techniques for implantable defibrillators American Heart Journal. ,vol. 127, pp. 1081- 1085 ,(1994) , 10.1016/0002-8703(94)90091-4
Soo G. Kim, Seymour Furman, Lawrence E. Waspe, Richard Brodman, John D. Fisher, Unipolar pacer artifacts induced failure of an automatic implantable cardioverter/defibrillator to detect ventricular fibrillation The American Journal of Cardiology. ,vol. 57, pp. 880- 881 ,(1986) , 10.1016/0002-9149(86)90634-X
Jodie L. Hurwitz, Bruce G. Hook, Belinda T. Flores, Francis E. Marchlinski, Importance of abortive shock capability with electrogram storage in cardioverter-defibrillator devices. Journal of the American College of Cardiology. ,vol. 21, pp. 895- 900 ,(1993) , 10.1016/0735-1097(93)90344-Z
FRANCIS E MARCHLINSKI, BELINDA T FLORES, ALFRED E BUXTON, W CLARK HARGROVE III, V PAUL ADDONIZIO, LARRY W STEPHENSON, ALDEN H HARKEN, JOHN U DOHERTY, E WAYNE GROGAN Jr, MARK E JOSEPHSON, The Automatic Implantable Cardioverter-Defibrillator: Efficacy, Complications, and Device Failures Annals of Internal Medicine. ,vol. 104, pp. 481- 488 ,(1986) , 10.7326/0003-4819-104-4-481
Peter D. Chapman, Paul Troup, The automatic implantable cardioverter-defibrillator: Evaluating suspected inappropriate shocks Journal of the American College of Cardiology. ,vol. 7, pp. 1075- 1078 ,(1986) , 10.1016/S0735-1097(86)80225-X
M. MIROWSKI, PHILIP R. REID, MORTON M. MOWER, LEVI WATKINS, EDWARD V. PLATIA, LAWRENCE S.C. GRIFFITH, THOMAS GUARNIERI, ANDRA THOMAS, JUAN M. JUANTEGUY, Clinical Performance of the Implantable Cardioverter‐Defibrillator Pacing and Clinical Electrophysiology. ,vol. 7, pp. 1345- 1350 ,(1984) , 10.1111/J.1540-8159.1984.TB05706.X
BERNARD DODINOT, JEAN-PHILIPPE GODENIR, ALVARO BARROS COSTA, CLAUDE ZELLER, MARK BROSCHART, Electronic Article Surveillance: A Possible Danger for Pacemaker Patients Pacing and Clinical Electrophysiology. ,vol. 16, pp. 46- 53 ,(1993) , 10.1111/J.1540-8159.1993.TB01534.X
Nancy J. Johnson, Francis E. Marchlinski, Arrhythmias induced by device antitachycardia therapy due to diagnostic nonspecificity. Journal of the American College of Cardiology. ,vol. 18, pp. 1418- 1425 ,(1991) , 10.1016/0735-1097(91)90566-R
FRANCIS E. MARCHLINSKI, CHARLES D. GOTTLIEB, BRIAN SARTER, JOHN FINKLE, BRUCE HOOK, DAVID CALLANS, DAVID SCHWARTZMAN, ICD data storage: value in arrhythmia management. Pacing and Clinical Electrophysiology. ,vol. 16, pp. 527- 534 ,(1993) , 10.1111/J.1540-8159.1993.TB01620.X