作者: B Daghman , M Piotin , C Mounayer , J Moret , L Spelle
DOI:
关键词:
摘要: BACKGROUND AND PURPOSE: The purpose of this study was to compare the volumetric results intracranial aneurysms obtained by calculation volume an ellipsoid with those 3D rotational angiography (3D-RA). METHODS: First, precision 3D-RA in assessment measurement aneurysm had be established. gave overestimation 4% 5.5% actual a spherical object. Then, 484 consecutive human were studied 3D-RA, allowing determination their volume. In meantime, dimensions (height and width) measured on pictures generated 3D-RA. volumes calculated (Vcalc) considering shape ellipsoidal. RESULTS: overestimated 15 ± 38% given Taking into account 10% margin error, 227 (47%) 44 34%, whereas 113 (23%) underestimated 25 12%. Only 144 (30%) within limits discrepancy. Concordance good for pericallosal basilar tip (mean 6 22% 8 27%, respectively). Conversely, there high discrepancy between posterior communicating artery 22 44%). CONCLUSION: based is relatively accurate aneurysms, probably owing or elliptic shape. formula not adequate irregularly shaped lesions, such as aneurysms.