Validation of a CFD Methodology for Positive Displacement LVAD Analysis Using PIV Data

作者: Richard B. Medvitz , Varun Reddy , Steve Deutsch , Keefe B. Manning , Eric G. Paterson

DOI: 10.1115/1.4000116

关键词:

摘要: Computational fluid dynamics (CFD) is used to asses the hydrodynamic performance of a positive displacement left ventricular assist device. The computational model uses implicit large eddy simulation direct resolution of the chamber compression and modeled valve closure to reproduce the in vitro results. The computations are validated through comparisons with experimental particle image velocimetry (PIV) data. Qualitative comparisons of flow patterns, velocity fields, and wall-shear rates demonstrate a high level of agreement between the computations and experiments. Quantitatively, the PIV and CFD show similar probed velocity histories, closely matching jet velocities and comparable wall-strain rates. Overall, it has been shown that CFD can provide detailed flow field and wall-strain rate data, which is important in evaluating blood pump performance.

参考文章(28)
Bill B. Daily, Timothy W. Pettitt, Salvatore P. Sutera, William S. Pierce, Pierce-Donachy pediatric VAD: Progress in development The Annals of Thoracic Surgery. ,vol. 61, pp. 437- 443 ,(1996) , 10.1016/0003-4975(95)00991-4
David E. Francischelli, John M. Tarbell, David B. Geselowitz, Local Blood Residence Times in the Penn State Artificial Heart Artificial Organs. ,vol. 15, pp. 218- 224 ,(1991) , 10.1111/J.1525-1594.1991.TB03042.X
Richard B. Medvitz, James W. Kreider, Keefe B. Manning, Arnold A. Fontaine, Steven Deutsch, Eric G. Paterson, Development and validation of a computational fluid dynamics methodology for simulation of pulsatile left ventricular assist devices. Asaio Journal. ,vol. 53, pp. 122- 131 ,(2007) , 10.1097/MAT.0B013E31802F37DD
P.C. Lu, H.C. Lai, J.S. Liu, A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow Journal of Biomechanics. ,vol. 34, pp. 1361- 1364 ,(2001) , 10.1016/S0021-9290(01)00084-7
Leslie A. Oley, Keefe B. Manning, Arnold A. Fontaine, Steven Deutsch, Off-design considerations of the 50cc Penn State Ventricular Assist Device. Artificial Organs. ,vol. 29, pp. 378- 386 ,(2005) , 10.1111/J.1525-1594.2005.29064.X
Sanjay M. Mehta, William J. Weiss, Alan J. Snyder, George A. Prophet, Walter E. Pae, Gerson Rosenberg, William S. Pierce, Testing of a 50 cc stroke volume completely implantable artificial heart: Expanding chronic mechanical circulatory support to women, adolescents, and small stature men Asaio Journal. ,vol. 46, pp. 779- 782 ,(2000) , 10.1097/00002480-200011000-00024
James W. Kreider, Keefe B. Manning, Leslie A. Oley, Arnold A. Fontaine, Steven Deutsch, The 50cc Penn State left ventricular assist device: a parametric study of valve orientation flow dynamics. Asaio Journal. ,vol. 52, pp. 123- 131 ,(2006) , 10.1097/01.MAT.0000199750.89636.77