作者: Carles Gomà , Bénédicte Hofstetter-Boillat , Sairos Safai , Sándor Vörös
DOI: 10.1088/0031-9155/60/8/3207
关键词:
摘要: This paper presents a method to experimentally validate the beam quality correction factors (kQ) tabulated in IAEA TRS-398 for proton beams and determine kQ of non-tabulated ionization chambers (based on already values). The is based exclusively ionometry it consists comparing reading two under same reference conditions Q 60Co. allows one ratio between chambers. In this work, 7 different chamber models were irradiated 60Co beams. For latter, both modulated (spread-out Bragg peak field) monoenergetic (pseudo-monoenergetic studied. beams, was found that experimental values obtained plane-parallel are consistent with TRS-398; whereas cylindrical not consistent?being higher than values. These results support suggestion (of previous publications) should be revised so effective point measurement taken into account when positioning at depth. all studied work each other?except IBA FC65-G, whose value 0.6% lower one. PTW Advanced Markus chamber, which TRS-398, 0.997???0.042 (k?=?2), chamber.