Parameter Modeling Analysis and Experimental Verification on Magnetic Shielding Cylinder of All-Optical Atomic Spin Magnetometer

作者: Hong Zhang , Sheng Zou , Xi-Yuan Chen , Wei Quan

DOI: 10.1155/2015/491746

关键词:

摘要: The ultrahigh sensitivity atomic spin magnetometer as the magnetic measurement sensor has received much concern. performance of shielding cylinder is one key factors constraining magnetometer’s sensitivity. In order to effectively improve performances shielding, parameter optimization models were established in this paper. Under condition changing only while others keeping constant, effects various parameters influencing axial coefficient comprehensively analyzed, and results showed that smaller innermost length, radius, radial spacing were, greater was, better could be obtained. According these actual needs, was optimally designed, then simulated via software Ansoft. simulation optimized had advantages small size, high performance, lager uniformity than nonoptimized one. residual magnetism below 0.1 nT, which 2~4 times lower

参考文章(20)
D. Sheng, S. Li, N. Dural, M. V. Romalis, Subfemtotesla scalar atomic magnetometry using multipass cells. Physical Review Letters. ,vol. 110, pp. 160802- ,(2013) , 10.1103/PHYSREVLETT.110.160802
I. K. Kominis, T. W. Kornack, J. C. Allred, M. V. Romalis, A subfemtotesla multichannel atomic magnetometer Nature. ,vol. 422, pp. 596- 599 ,(2003) , 10.1038/NATURE01484
Jiancheng Fang, Tao Wang, Wei Quan, Heng Yuan, Hong Zhang, Yang Li, Sheng Zou, In situ magnetic compensation for potassium spin-exchange relaxation-free magnetometer considering probe beam pumping effect. Review of Scientific Instruments. ,vol. 85, pp. 063108- ,(2014) , 10.1063/1.4881685
S. Appelt, A. Ben-Amar Baranga, C. J. Erickson, M. V. Romalis, A. R. Young, W. Happer, Theory of spin-exchange optical pumping of 3 He and 129 Xe Physical Review A. ,vol. 58, pp. 1412- 1439 ,(1998) , 10.1103/PHYSREVA.58.1412
Gorazd Planinšič, Shielding of Low-Frequency Magnetic Interference in Weak-Field MRI by a Single-Layer Cylindrical Coil Journal of Magnetic Resonance. ,vol. 126, pp. 30- 38 ,(1997) , 10.1006/JMRE.1997.1144
L. Xuan, H. Dong, C. Zhuo, Compact multi-layer magnetic shields for chip-scale atomic devices nano/micro engineered and molecular systems. pp. 649- 652 ,(2013) , 10.1109/NEMS.2013.6559813
V. Altuzar, S. A. Tomás, O. Zelaya-Angel, F. Sánchez-Sinencio, J. L. Arriaga, Atmospheric pollution profiles in Mexico City in two different seasons Review of Scientific Instruments. ,vol. 74, pp. 500- 502 ,(2003) , 10.1063/1.1512768
T. W. Kornack, M. V. Romalis, Dynamics of two overlapping spin ensembles interacting by spin exchange. Physical Review Letters. ,vol. 89, pp. 253002- ,(2002) , 10.1103/PHYSREVLETT.89.253002
Dahai Ren, Lingqi Wu, Meizhi Yan, Mingyang Cui, Zheng You, Muzhi Hu, Design and Analyses of a MEMS Based Resonant Magnetometer Sensors. ,vol. 9, pp. 6951- 6966 ,(2009) , 10.3390/S90906951
T. H. Sander, J. Preusser, R. Mhaskar, J. Kitching, L. Trahms, S. Knappe, Magnetoencephalography with a chip-scale atomic magnetometer Biomedical Optics Express. ,vol. 3, pp. 981- 990 ,(2012) , 10.1364/BOE.3.000981