Digital chest radiography with a solid-state flat-panel x-ray detector: contrast-detail evaluation with processed images printed on film hard copy.

作者: Harrell G. Chotas , Carl E. Ravin

DOI: 10.1148/RADIOLOGY.218.3.R01MR27679

关键词:

摘要: PURPOSE: To evaluate and compare human observer performance in a contrast-detail test by using postprocessed hard-copy images from digital chest radiography system conventional screen-film radiographs. MATERIALS AND METHODS: The is based on large-area flat-panel x-ray detector with structured cesium iodide scintillator layer an amorphous silicon thin-film transistor array for image readout. Images of phantom were acquired at two exposure levels standard thoracic systems the matched dose. By processed postprocessing techniques, four-alternative forced-choice perception study was performed, number detectable signals (disk-shaped objects 0.3–4.0 mm diameter) determined each type. RESULTS: On average, observers detected more than radiographs all diameters up to...

参考文章(5)
David L. Ergun, Lois Rutz, Dean A. Hinshaw, Hartwig Blume, Dwayne C. Clark, James T. Dobbins, DQE(f ) of four generations of computed radiography acquisition devices Medical Physics. ,vol. 22, pp. 1581- 1593 ,(1995) , 10.1118/1.597627
Paul R. Granfors, Performance characteristics of an amorphous silicon flat-panel x-ray imaging detector Proceedings of SPIE, the International Society for Optical Engineering. ,vol. 3659, pp. 480- 490 ,(1999) , 10.1117/12.349525
Ulrich Neitzel, Ingo Maack, Susanne Günther-Kohfahl, Image quality of a digital chest radiography system based on a selenium detector Medical Physics. ,vol. 21, pp. 509- 516 ,(1994) , 10.1118/1.597389
P. Burt, E. Adelson, The Laplacian Pyramid as a Compact Image Code IEEE Transactions on Communications. ,vol. 31, pp. 671- 679 ,(1983) , 10.1109/TCOM.1983.1095851