Automatic axial range selection on the CT localizer to remove unused SPECT/CT slices: Application to serial imaging in radiopharmaceutical therapy

作者: Yazdan Salimi , Zahra Mansouri , Azadeh Akhavanallaf , Isaac Shiri , Eliluane Teixeira

DOI:

关键词:

摘要: P1102Introduction: Radiopharmaceutical therapy (RPT) has shown promising outcomes in the palliative treatment of cancer patients. Personalized dosimetry in theranostics is essential to estimate the radiation doses delivered to tumors and organs at risk. In this regard, two or preferably three quantitative SPECT images are required to yield the time-integrated activity (TIA) maps obtained by fitting a curve to model the pharmacokinetics of the radiotracer in the patient’s body. Registration of SPECT/CT images among the time points is necessary to extract the time-activity curve (TAC) required for both organ-level and voxel-wise dosimetry. As such, these images should be acquired by covering the same axial field-of-view (FOV) covered by commonly used ~40 cm/bed SPECT/CT cameras. Yet, in reality, there may be mismatches (figure 1a). We evaluated this mismatch in patients treated by Lu-177 RPT and developed …