Cardiac SPECT radiomic features repeatability and reproducibility: A multi-scanner phantom study

作者: Mohammad Edalat-Javid , Isaac Shiri , Ghasem Hajianfar , Hamid Abdollahi , Hossein Arabi

DOI:

关键词:

摘要: BackgroundThe aim of this work was to assess the robustness of cardiac SPECT radiomic features against changes in imaging settings, including acquisition, and reconstruction parameters.MethodsFour commercial SPECT and SPECT/CT cameras were used to acquire images of a static cardiac phantom mimicking typical myorcardial perfusion imaging using 185 MBq of 99mTc. The effects of different image acquisition and reconstruction parameters, including number of views, view matrix size, attenuation correction, as well as image reconstruction related parameters (algorithm, number of iterations, number of subsets, type of post-reconstruction filter, and its associated parameters, including filter order and cut-off frequency) were studied. In total, 5,063 transverse views were reconstructed by varying the aforementioned factors. Eighty-seven radiomic features including first-, second-, and high-order textures were …