Evaluation of the portable pressure transducer technique for measuring field tensiometers [soil moisture]

作者: HP Cresswell

DOI:

关键词:

摘要: The use of a portable pressure transducer system which couples to tensiometers via a syringe needle and septum stopper was assessed. Changes that occurred to pressure within the tensiometer on insertion of the needle were significant and up to 120 min was sometimes required before complete re-equilibration to the original tensiometer pressure was achieved. Transducer readings should be made after allowing time for complete re-equilibration of the tensiometer following insertion of the needle. However, for practical field measurement of low matric potentials, it will be necessary to read the portable transducer before full re-equilibration and correct the reading based on previously determined re-equilibration patterns. These re-equilibration patterns should be specific to each tensiometer and should cover a range of matric potentials. It is also important that the air bubble below the septum stopper be maintained constant for each reading, and that the septum stoppers be replaced regularly. With careful use, these systems are convenient for monitoring large numbers of tensiometers spaced apart in the field.

参考文章(0)