Sterilization effects on nitric oxide-releasing glucose sensors

作者: Taron M Bradshaw , Courtney R Johnson , Christopher A Broberg , Darci E Anderson , Mark H Schoenfisch

DOI:

关键词:

摘要: Nitric oxide (NO) release from S-nitrosothiol-modified mesoporous silica nanoparticles imbedded in the diffusion limiting layer of a glucose sensor has been demonstrated as an effective strategy for mitigating the foreign body response common to sensor implantation, resulting in improved analytical performance. With respect to potential clinical translation of this approach, the effects of sterilization on NO-releasing biosensors require careful evaluation, as NO donor chemistry is sensitive to temperature and environment. Herein, we evaluated the influence of multiple sterilization methods on 1) sterilization success; 2) NO payload; and 3) sensor performance to establish the commercialization potential of NO-releasing glucose sensors. Sensors were treated with ethylene oxide gas, the most common sterilization method for intricate medical devices, which led to undesirable (i.e., premature) release of NO. To reduce …