Evaluation of three different specimen types (serum, plasma lithium heparin and serum gel separator) for analysis of certain analytes: clinical significance of differences in results and efficiency in use.

作者: Myra P. O'Keane , Sean K. Cunningham

DOI: 10.1515/CCLM.2006.099

关键词: Serum plasmaAnalyteHeparinPhentytoinClinical significanceChemistryChromatographyPotassiumCreatinineBlood plasma

摘要: Background There is a lack of consensus regarding the most appropriate specimen type for analysis many biochemistry analytes. The aim this study was to compare renal and lipid analyte profiles phenytoin values in plain serum (S), gel (G) plasma (lithium heparin, P) tubes investigate stability these analytes after prolonged contact with cells or at room temperature (RT, 20 degrees C) as aliquoted stored 4 C. Methods Primary specimens were centrifuged once, maintained RT analysed within 2 h (T(0)) 24 (T(24)) 48 (T(48)). For assessment C, two cell-free aliquots separated from each primary C then T(24) T(48). Differences concentrations between T(0) following storage (at T(48)) evaluated both statistical clinical significance. Results At all analytes, except potassium, demonstrated equivalence serum, tubes. Potassium creatinine more stable than serum/plasma In contrast, phentytoin up T(48) RT, but showed progressive clinically significant decrease concentration RT. All CO(2) when Conclusions We concluded that tube has advantages over measurement investigated study, exception phenytoin. practice, demonstrate enhanced reduce need aliquot specimens, greater protection against possible contamination. Further investigation would be required evaluate broader range

参考文章(35)
Jack H. Ladenson, Lii-Mei B. Tsai, J. M. Michael, G. Kessler, J. Heinz Joist, Serum versus heparinized plasma for eighteen common chemistry tests: is serum the appropriate specimen? American Journal of Clinical Pathology. ,vol. 62, pp. 545- 552 ,(1974) , 10.1093/AJCP/62.4.545
Robert D. Herr, Terry Swanson, Serum Bicarbonate Declines with Sample Size in Vacutainer Tubes American Journal of Clinical Pathology. ,vol. 97, pp. 213- 216 ,(1992) , 10.1093/AJCP/97.2.213
Gifford Lum, S. Raymond Gambino, A Comparison of Serum versus Heparinized Plasma for Routine Chemistry Tests American Journal of Clinical Pathology. ,vol. 61, pp. 108- 113 ,(1974) , 10.1093/AJCP/61.1.108
[edited by]Carl A. Burtis, Edward A. Ashwood, Tietz textbook of clinical chemistry : W.B.Saunders. ,(2015)
Albert Hanok, Jeremiah Kuo, The Stability of a Reconstituted Serum for the Assay of Fifteen Chemical Constituents Clinical Chemistry. ,vol. 14, pp. 58- 69 ,(1968) , 10.1093/CLINCHEM/14.1.58
J C Smith, S Lewis, J Holbrook, K Seidel, A Rose, Effect of heparin and citrate on measured concentrations of various analytes in plasma. Clinical Chemistry. ,vol. 33, pp. 814- 816 ,(1987) , 10.1093/CLINCHEM/33.6.814
Thomas R. Koch, Gennady Platoff, Suitability of collection tubes with separator gels for therapeutic drug monitoring. Therapeutic Drug Monitoring. ,vol. 12, pp. 277- 280 ,(1990) , 10.1097/00007691-199005000-00011
Michael Heins, Wolfgang Heil, Wolfgang Withold, Storage of serum or whole blood samples? Effects of time and temperature on 22 serum analytes. Clinical Chemistry and Laboratory Medicine. ,vol. 33, pp. 231- 238 ,(1995) , 10.1515/CCLM.1995.33.4.231
John G Mohler, Pamela A. Mohler, Pharm D, Rosalind G Pallivathucal, Failure of the serum CO2 determined by automation to estimate the plasma bicarbonate. Scandinavian Journal of Clinical & Laboratory Investigation. ,vol. 47, pp. 61- 67 ,(1987) , 10.1080/00365518709168174