作者: Johnny M Khoury , Toufic Farah , Christelle P El-Haibi , Baha N Noureddin
DOI: 10.3928/1081-597X-20070901-11
关键词: Biomedical engineering 、 Excimer 、 Ablation 、 Mitomycin C 、 Delivery system 、 Corneal light shield 、 Coefficient of variation 、 Filter paper 、 Optics 、 Photorefractive keratectomy 、 Materials science
摘要: PURPOSE To develop a simple, reproducible method of applying intraoperative mitomycin C (MMC) in excimer surface ablation surgery. METHODS A two-part protocol was developed to study several properties corneal light shields. Part tested the amount MMC (0.2 mg/mL) absorbed, expansion dimension attained, and released filter paper. B examined vitro release photorefractive keratectomy (PRK)-treated enucleated pig eye. RESULTS Mean absorbed by shields 63.7 +/- 5.0 mg (range: 57.4 72.8 mg); coefficient variability 7.8%. Expansion diameter thickness were constant at 9.0 mm 1.1 mm, respectively. solution paper 55.4 3.9 51.3 63.5 7.1%. transferred PRK-treated eye 2.3 0.7 3.5 33%. CONCLUSIONS Mitomycin contact area, each shield reproducible. Uniform dimensions theoretically provide uniform distribution MMC. This may allow standardization application ablation.