Ultramicrobiosensors for monitoring of neurotransmitters.

作者: EIICHI TAMIYA , ISAO KARUBE

DOI: 10.1111/J.1749-6632.1992.TB32690.X

关键词: PotassiumCalibration curveChemistryElectrochemistryCerebellar cortexHydrogen peroxideElectrodeCholine oxidaseMembraneAnalytical chemistryChromatography

摘要: Carbon fiber electrodes are used to construct ultramicrobiosensors with 7-15 microns diameter. Electrochemical operations for preelectrolysis and measuring were examined sensitive determination of hydrogen peroxide. Determination limit was 0.1 microM Reproducible peroxide is possible even in samples containing albumin protein. A micro-acetylcholine sensor fabricated by immobilizing acetylcholine esterase choline oxidase on the carbon entrapment PVA-SbQ. This gave a linear calibration plot range from 1.0 mM correlation coefficient 0.9842. micro-glutamate consisted platinized disk electrode modified immobilized glutamate membrane. 2 1.2 mM. Release cerebellar cortex detected after potassium stimulation.

参考文章(6)
Taizo Nakazato, Akitane Akiyama, In vivo voltammetric study of 6-hydroxydopamine-induced neuronal degradation. Journal of Neurochemistry. ,vol. 51, pp. 1007- 1013 ,(1988) , 10.1111/J.1471-4159.1988.TB03061.X
Koichi Aoki, Minoru Ishida, Koichi Tokuda, Kiyoshi Hasebe, Electrode kinetics of the oxidation of hydrogen peroxide at pretreated glassy carbon and carbon fiber electrodes Journal of Electroanalytical Chemistry and Interfacial Electrochemistry. ,vol. 251, pp. 63- 71 ,(1988) , 10.1016/0022-0728(88)80385-1
Eiichi Tamiya, Isao Karube, None, Micro-biosensors for clinical analyses Sensors and Actuators. ,vol. 15, pp. 199- 207 ,(1988) , 10.1016/0250-6874(88)87009-4
Akitane. Akiyama, Takeshi. Kato, Kayoko. Ishii, Eiichi. Yasuda, In vitro measurement of dopamine concentration with carbon fiber electrode Analytical Chemistry. ,vol. 57, pp. 1518- 1522 ,(1985) , 10.1021/AC00285A006
Eiichi Tamiya, Youko Sugiura, Eriberta Nepomuceno Navera, Shingo Mizoshita, Kenji Nakajima, Akitane Akiyama, Isao Karube, Ultramicro acetylcholine sensor based on an enzyme-modified carbon fibre electrode Analytica Chimica Acta. ,vol. 251, pp. 129- 134 ,(1991) , 10.1016/0003-2670(91)87125-Q