MRI and implanted medical devices: basic interactions with an emphasis on heating

作者: J.A. Nyenhuis , Sung-Min Park , R. Kamondetdacha , A. Amjad , F.G. Shellock

DOI: 10.1109/TDMR.2005.859033

关键词: Emphasis (telecommunications)Lead (electronics)ElectrodeMagnetic resonance imagingMaterials scienceRadio frequencyMagnetic fieldRF power amplifierBiomedical engineeringAudio frequency

摘要: … The power deposition near the ends of a lead wire can be … Resonant circuits with inductance and capacitance were found … of lead resistance and other approximations in the model. …

参考文章(50)
M. B. Dragović, A. R. Djordjević, Branko D. Popović, Analysis and synthesis of wire-antennas ,(1982)
P. So, K. Caputa, M.A. Stuchly, Peripheral nerve stimulation by gradient switching fields in MRI international conference of the ieee engineering in medicine and biology society. ,vol. 4, pp. 3371- 3374 ,(2003) , 10.1109/IEMBS.2003.1280981
Chris D Smith, Alexander V Kildishev, John A Nyenhuis, Kirk S Foster, Joe D Bourland, None, Interactions of magnetic resonance imaging radio frequency magnetic fields with elongated medical implants Journal of Applied Physics. ,vol. 87, pp. 6188- 6190 ,(2000) , 10.1063/1.372651
T. Whit Athey, A model of the temperature rise in the head due to magnetic resonance imaging procedures. Magnetic Resonance in Medicine. ,vol. 9, pp. 177- 184 ,(1989) , 10.1002/MRM.1910090204
Barrie Condon, Donald M. Hadley, Potential MR Hazard to Patients With Metallic Heart Valves: The Lenz Effect Journal of Magnetic Resonance Imaging. ,vol. 12, pp. 171- 176 ,(2000) , 10.1002/1522-2586(200007)12:1<171::AID-JMRI19>3.0.CO;2-W
Wolfgang R. Nitz, Arnulf Oppelt, Wolfgang Renz, Christoph Manke, Markus Lenhart, Johann Link, On the heating of linear conductive structures as guide wires and catheters in interventional MRI. Journal of Magnetic Resonance Imaging. ,vol. 13, pp. 105- 114 ,(2001) , 10.1002/1522-2586(200101)13:1<105::AID-JMRI1016>3.0.CO;2-0
Kenneth B. Baker, John A. Nyenhuis, Greg Hrdlicka, Ali R. Rezai, Jean A. Tkach, Frank G. Shellock, Neurostimulation systems: assessment of magnetic field interactions associated with 1.5- and 3-Tesla MR systems. Journal of Magnetic Resonance Imaging. ,vol. 21, pp. 72- 77 ,(2005) , 10.1002/JMRI.20224
Richard C. Hartwell, Frank G. Shellock, MRI of cervical fixation devices: Sensation of heating caused by vibration of metallic components Journal of Magnetic Resonance Imaging. ,vol. 7, pp. 771- 772 ,(1997) , 10.1002/JMRI.1880070427
Ali R. Rezai, Daniel Finelli, John A. Nyenhuis, Greg Hrdlicka, Jean Tkach, Ashwini Sharan, Paul Rugieri, Paul H. Stypulkowski, Frank G. Shellock, Neurostimulation systems for deep brain stimulation: in vitro evaluation of magnetic resonance imaging-related heating at 1.5 tesla. Journal of Magnetic Resonance Imaging. ,vol. 15, pp. 241- 250 ,(2002) , 10.1002/JMRI.10069