Towards true unipolar ECG recording without the Wilson central terminal (preliminary results).

作者: G D Gargiulo , A L McEwan , P Bifulco , M Cesarelli , C Jin

DOI: 10.1088/0967-3334/34/9/991

关键词: SoftwareElectrical engineeringContinuous monitoringWearable computerTerminal (electronics)Computer scienceGold standard (test)WidebandPower (physics)Lead (electronics)

摘要: We present an innovative bio-potential front-end capable of recording true unipolar ECG leads for the first time without making use Wilson central terminal. In addition to convenience in applications such as continuous monitoring and rapid diagnosis, information recordings may yield unique diagnostic it avoids need essentially subtract data or make averaging effect imposed from The system also allows direct, real-time software calculation signals corresponding standard which achieve correlations excess 92% with a gold during parallel vivo recording. addition, implemented circuit is wideband (0.05?1000?Hz), compatible (Ag/AgCl) electrodes, dry (paste-less) textile electrodes. low power, requiring less than 50 mW (when powered at 12?V) per lead (two channels required). It therefore well suited wearable, long-term applications.

参考文章(16)
Oscar Aquilina, A brief history of cardiac pacing. Images in paediatric cardiology. ,vol. 8, pp. 17- 81 ,(2006)
Metin Akay, Wiley encyclopedia of biomedical engineering Wiley-Interscience. ,(2006)
John D. Enderle, Joseph D. Bronzino, Susan M. Blanchard, Introduction to biomedical engineering Academic Press. ,(2000)
Paolo Bifulco, Gaetano Gargiulo, Mario Cesarelli, Alistair McEwan, Craig Jin, Andre van Schaik, Wearable dry sensors with bluetooth connection for use in remote patient monitoring systems. Studies in health technology and informatics. ,vol. 161, pp. 57- 65 ,(2010) , 10.3233/978-1-60750-659-1-57
Gaetano D Gargiulo, Alistair L McEwan, Paolo Bifulco, Mario Cesarelli, Craig Jin, Jonathan Tapson, Aravinda Thiagalingam, André van Schaik, Towards true unipolar bio-potential recording: a preliminary result for ECG Physiological Measurement. ,vol. 34, ,(2013) , 10.1088/0967-3334/34/1/N1
Gaetano Gargiulo, Paolo Bifulco, Mario Cesarelli, Mariano Ruffo, Maria Romano, Rafael A Calvo, Craig Jin, André van Schaik, An ultra-high input impedance ECG amplifier for long-term monitoring of athletes Medical Devices : Evidence and Research. ,vol. 3, pp. 1- 9 ,(2010) , 10.2147/MDER.S9321
I.K. Duskalov, I.A. Dotsinsky, I.I. Christov, Developments in ECG acquisition, preprocessing, parameter measurement, and recording IEEE Engineering in Medicine and Biology Magazine. ,vol. 17, pp. 50- 58 ,(1998) , 10.1109/51.664031
R J Prance, A Debray, T D Clark, H Prance, M Nock, C J Harland, A J Clippingdale, An ultra-low-noise electrical-potential probe for human-body scanning Measurement Science and Technology. ,vol. 11, pp. 291- 297 ,(2000) , 10.1088/0957-0233/11/3/318
A Searle, L Kirkup, A direct comparison of wet, dry and insulating bioelectric recording electrodes. Physiological Measurement. ,vol. 21, pp. 271- 283 ,(2000) , 10.1088/0967-3334/21/2/307