作者: M. Tibben , J. Rademaker-Lakhai , J. Rice , D. Stewart , J. Schellens
DOI: 10.1007/S00216-002-1340-9
关键词: N-(2-Hydroxypropyl) methacrylamide 、 Methacrylamide 、 Detection limit 、 Platinum 、 Atomic absorption spectroscopy 、 Mass spectrometry 、 Chromatography 、 Quantitative analysis (chemistry) 、 Calibration curve 、 Chemistry
摘要: AP5280 is an N-(2-hydroxypropyl)methacrylamide (HPMA) copolymer to which are linked tetrapeptide side chains containing bioactive platinum complexes at their C-terminal sides. We have developed and validated a rapid sensitive analytical assay for the determination of total concentrations in plasma, free origin plasma ultrafiltrate (PUF), subjects dosed with AP5280. The levels were determined by use graphite-furnace atomic-absorption spectrometry (GF-AAS) Zeeman correction after appropriate dilution sample plasma–hydrochloric acid 0.2 mol L–1 (1:5) as diluent. limit quantitation this 0.25 µmol plasma. Linear calibration curves obtained over concentration range 0.25–5.0 L–1. Accuracy was between 87.7% 104.2% precision 15.3% lowest less than 14% all other tested. thus accordance generally accepted criteria methods. Analysis samples from patients receiving demonstrated applicability described assay.