作者: Rita M. Bruver , Mark L. Salkie
DOI: 10.1016/S0009-9120(78)90126-1
关键词: Nephelometer 、 Blood testing 、 Blood proteins 、 Analytical chemistry 、 Radial immunodiffusion 、 Nephelometry 、 Medicine 、 Chromatography
摘要: The Laser Nephelometer PDQ™ (Hyland Division, Travenol Laboratories Inc.) and the Abbott Bichromatic Analyser 100 (Abbott Laboratories) were compared with a radial immunodiffusion method. Seventy-eight serum specimens collected during routine blood testing aliquoted quantitated by three procedures described. nephelometric system was used as described Hyland in their instructions accompanying LAS-R Nephelometric test kits. ABA-100 filter unit transmitting at 340 650 nm water bath temperature of 30°C. correlated well RID giving correlation coefficient varying from 0.94 for IgA to 0.79 C3. It not possible quantitate turbidometric method using ABA-100. Results obtained other proteins satisfactory varied 0.88 IgG 0.90 procedure use takes 5 hours technologist time 16-hour incubation period produce results IgG, IgM on nine patient specimens. Using obtain same 20 took two hours. Nephelometry, therefore, appears be alternative comparable precision great saving time.