作者: H Avram , H Chen , L Kaufman
DOI: 10.1148/RADIOLOGY.189.3.8234721
关键词: Mr imaging 、 Signal 、 Image processing 、 High-pass filter 、 Medicine 、 Nuclear magnetic resonance 、 Degradation (telecommunications) 、 Low-pass filter 、 Optics 、 k-space 、 Reduction (complexity)
摘要: PURPOSE: To evaluate the use of conjugation with hybrid magnetic resonance imaging techniques in order to reduce artifacts from inconsistencies introduced by T2 decay. MATERIALS AND METHODS: Two phantoms containing either oil or water were used study edge spread function (ESF) and signal-to-noise ratio (S/N) for three acquisition strategies (low pass, high ramp) which a full set was acquired partial conjugation. RESULTS: With half echoes providing coverage same amount k space, signal reduced factor at least e. Conjugation improved ESF degradation, sequence efficiency increased over 50%, compensated loss S/N. CONCLUSION: The time saved can be nearly double number sections given repetition (TR) obtain two simultaneous instead one, reduction artifacts. Alternatively, halved TR ...