A scalable and modular lab-on-a-chip genetic analysis instrument

作者: G. V. Kaigala , M. Behnam , A. C. E. Bidulock , C. Bargen , R. W. Johnstone

DOI: 10.1039/B925111A

关键词: Orders of magnitude (temperature)Sensitivity (control systems)Instrumentation (computer programming)ScalabilityModular designComputer scienceLab-on-a-chipComponent (UML)Embedded systemSystem on a chip

摘要: We demonstrate a new and extremely inexpensive, multipurpose desktop system for operating lab-on-a-chip (LOC) devices. The provides all of the infrastructure necessary genetic amplification analysis, with orders magnitude improvement in performance over our previous work. A modular design enables high levels integration while allowing scalability to lower cost smaller size. component this is ca. $600, yet it could support many diagnostic applications. an implementation via polymerase chain reaction (PCR), analysis using capillary electrophoresis (CE). PCR able amplify from single or several copies target DNA CE (e.g. sensitivity) comparable that commercial photomultiplier-based confocal lab-on-chip instrumentation. believe demonstrates need no longer be barrier wide-spread application LOC technologies healthcare beyond.

参考文章(36)
S. Karlin, S. F. Altschul, Applications and statistics for multiple high-scoring segments in molecular sequences. Proceedings of the National Academy of Sciences of the United States of America. ,vol. 90, pp. 5873- 5877 ,(1993) , 10.1073/PNAS.90.12.5873
K. Kolari, R. Satokari, K. Kataja, J. Stenman, A. Hokkanen, Real-time analysis of PCR inhibition on microfluidic materials Sensors and Actuators B: Chemical. ,vol. 128, pp. 442- 449 ,(2008) , 10.1016/J.SNB.2007.06.034
Jeong Ah Kim, Ji Youn Lee, Shimyoung Seong, Seung Hwan Cha, Seung Hwan Lee, Jae Jeong Kim, Tai Hyun Park, Fabrication and characterization of a PDMS–glass hybrid continuous-flow PCR chip Biochemical Engineering Journal. ,vol. 29, pp. 91- 97 ,(2006) , 10.1016/J.BEJ.2005.02.032
E. T. Lagally, H. T. Soh, Integrated Genetic Analysis Microsystems Critical Reviews in Solid State and Materials Sciences. ,vol. 30, pp. 207- 233 ,(2005) , 10.1080/10408430500332149
Govind V. Kaigala, Ryan J. Huskins, Jutta Preiksaitis, Xiao-Li Pang, Linda M. Pilarski, Christopher J. Backhouse, Automated screening using microfluidic chip-based PCR and product detection to assess risk of BK virus-associated nephropathy in renal transplant recipients. Electrophoresis. ,vol. 27, pp. 3753- 3763 ,(2006) , 10.1002/ELPS.200600061
Daniel MacDougall, Warren B Crummett, None, Guidelines for data acquisition and data quality evaluation in environmental chemistry Analytical Chemistry. ,vol. 52, pp. 2242- 2249 ,(1980) , 10.1021/AC50064A004
Jing-Lin Fu, Qun Fang, Ting Zhang, Xin-Hua Jin, Zhao-Lun Fang, Laser-Induced Fluorescence Detection System for Microfluidic Chips Based on an Orthogonal Optical Arrangement Analytical Chemistry. ,vol. 78, pp. 3827- 3834 ,(2006) , 10.1021/AC060153Q
G.V. Kaigala, M. Behnam, C. Bliss, M. Khorasani, S. Ho, J.N. McMullin, D.G. Elliott, C.J. Backhouse, Inexpensive, universal serial bus-powered and fully portable lab-on-a-chip-based capillary electrophoresis instrument Iet Nanobiotechnology. ,vol. 3, pp. 1- 7 ,(2009) , 10.1049/IET-NBT:20080005
Robert J. Meagher, Anson V. Hatch, Ronald F. Renzi, Anup K. Singh, An integrated microfluidic platform for sensitive and rapid detection of biological toxins Lab on a Chip. ,vol. 8, pp. 2046- 2053 ,(2008) , 10.1039/B815152K