作者: Vanessa Pincham , Jon Hyett , Karen Pollard , Philip Schluter , Andrew McLennan
DOI: 10.1002/AJUM.12000
关键词: Operator training 、 Quality assurance 、 Doppler measurements 、 Tricuspid valve 、 Nuclear medicine 、 Ductus venosus 、 Surgery 、 Medicine 、 Intraclass correlation 、 Doppler effect 、 Reference range
摘要: Objective To develop reference ranges for the ductus venosus (DV) and tricuspid valve (TV) waveforms at 11–14 weeks define auditable standards to assess operator measurement performance. Materials methods A single prospectively obtained quantitative measurements of a number DV TV Doppler indices medians 90% intervals (RIs). Measurement agreement studies between two experienced operators were also performed. The bias three additional was subsequently assessed using newly defined standards. Results Doppler in 292 patients 321 (TV). Reference constructed pulsatility index veins (PIV) which did not change over gestation (mean 1.06; RI 0.86–1.23) E–A ratio range positively correlated with gestation. on 30 showed within-operator almost perfect PIV (ICC, intraclass correlation 0.82–0.86) strong (ICC 0.68–0.78) while between-operator good both 0.46 both). Discussion Development local enabled definition that can be utilised assessment training ongoing quality assurance. Measurements by found reproducible.