Accuracy of oxygen measurements in T2 -(line width) EPR oximetry

作者: Tatyana I. Smirnova , Alex I. Smirnov , Robert B. Clarkson , R. Linn Belford

DOI: 10.1002/MRM.1910330610

关键词: ChemistryLine (formation)OxygenHomogeneous broadeningSpectral lineAnalytical chemistryElectron paramagnetic resonanceProtonLimiting oxygen concentrationNitroxide mediated radical polymerization

摘要: EPR oximetry is used for in vivo and vitro measurements of oxygen biological systems, including experimental animals. The accuracy T2 (line width) significantly improved if least-squares simulation to extract the line width parameters. effect on spectra nitroxide solutions aqueous suspensions fusinite can be described as an additional homogeneous broadening that modifies spectrum oxygen-free probe. This allows one use a one-parameter model most cases. simulations were carried out with fast-convolution algorithm followed by Levenberg-Marquardt optimization. validity error estimates provided this method was tested sets taken under common conditions. It shown (T2) determined not only probe sensitivity (rate change versus concentration), but also signal-to-noise ratio, inhomogeneous contribution shape (e.g., unresolved proton superhyperfine structure), spectral window. compared two radicals different structure fusinite.

参考文章(26)
H. M. Swartz, S. Boyer, D. Brown, K. Chang, P. Gast, J. F. Glockner, H. Hu, K. J. Liu, M. Moussavi, M. Nilges, S. W. Norby, A. Smirnov, N. Vahidi, T. Walczak, M. Wu, R. B. Clarkson, The use of EPR for the measurement of the concentration of oxygen in vivo in tissues under physiologically pertinent conditions and concentrations. Advances in Experimental Medicine and Biology. ,vol. 317, pp. 221- 228 ,(1992) , 10.1007/978-1-4615-3428-0_23
J.C. Evans, P.H. Morgan, R.H. Renaud, Simulation of electron spin resonance spectra by fast fourier transform Analytica Chimica Acta. ,vol. 103, pp. 175- 187 ,(1978) , 10.1016/S0003-2670(01)84037-X
H.J. Halpern, M. Peric, C. Yu, B.L. Bales, Rapid Quantitation of Parameters from Inhomogeneously Broadened EPR Spectra Journal of Magnetic Resonance, Series A. ,vol. 103, pp. 13- 22 ,(1993) , 10.1006/JMRA.1993.1125
Garry R. Buettner, Kitonga P. Kiminyo, Optimal EPR detection of weak nitroxide spin adduct and ascorbyl free radical signals. Journal of Biochemical and Biophysical Methods. ,vol. 24, pp. 147- 151 ,(1992) , 10.1016/0165-022X(92)90054-E
G. M. Rosen, H. J. Halpern, L. A. Brunsting, D. P. Spencer, K. E. Strauss, M. K. Bowman, A. S. Wechsler, Direct measurement of nitroxide pharmacokinetics in isolated hearts situated in a low-frequency electron spin resonance spectrometer: implications for spin trapping and in vivo oxymetry. Proceedings of the National Academy of Sciences of the United States of America. ,vol. 85, pp. 7772- 7776 ,(1988) , 10.1073/PNAS.85.20.7772
Alex I. Smirnov, Shong-Wan Norby, R. B. Clarkson, Ted Walczak, Harold M. Swartz, Simultaneous multi-site EPR spectroscopy in vivo Magnetic Resonance in Medicine. ,vol. 30, pp. 213- 220 ,(1993) , 10.1002/MRM.1910300210
Harold M. Swartz, Ke Jian Liu, Fuminori Goda, Tadeusz Walczak, India ink : a potential clinically applicable EPR oximetry probe Magnetic Resonance in Medicine. ,vol. 31, pp. 229- 232 ,(1994) , 10.1002/MRM.1910310218
C. S. Lai, L. E. Hopwood, J. S. Hyde, S. Lukiewicz, ESR studies of O2 uptake by Chinese hamster ovary cells during the cell cycle Proceedings of the National Academy of Sciences of the United States of America. ,vol. 79, pp. 1166- 1170 ,(1982) , 10.1073/PNAS.79.4.1166
Alex I. Smirnov, Shong-Wan Norby, James A. Weyhenmeyer, R.B. Clarkson, The effect of temperature on the respiration of cultured neural cells as studied by a novel electron paramagnetic resonance technique Biochimica et Biophysica Acta. ,vol. 1200, pp. 205- 214 ,(1994) , 10.1016/0304-4165(94)90137-6
A.I. Smirnov, R.L. Belford, Rapid Quantitation from Inhomogeneously Broadened EPR Spectra by a Fast Convolution Algorithm Journal of Magnetic Resonance, Series A. ,vol. 113, pp. 65- 73 ,(1995) , 10.1006/JMRA.1995.1057