Standardized Evaluation of 2D-3D Registration

作者: Everine B. van de Kraats , Graeme P. Penney , Dejan Tomaževič , Theo van Walsum , Wiro J. Niessen

DOI: 10.1007/978-3-540-30135-6_70

关键词:

摘要: In the past few years a number of 2D-3D registration algorithms have been introduced. However, these methods not directly compared or only work for specific applications. Understanding and evaluating their performance is therefore an open important issue. To address this challenge we introduce standard evaluation method, which can be used all types different Our method uses geometry 3D Rotational X-ray (3DRX) imaging system in combination with 3D-3D attaining highly accurate ground truth 2D multiple to MR/CT/3DRX registration. The data transformations will made available on Internet. Furthermore, propose starting positions failure criteria allow future researchers compare methods. As illustration, proposed has evaluate two techniques, viz. gradient-based intensity-based spinal

参考文章(8)
Everine B. van de Kraats, Theo van Walsum, Jorrit-Jan Verlaan, F. Cumhur Öner, Max A. Viergever, Wiro J. Niessen, Noninvasive Magnetic Resonance to Three-Dimensional Rotational X-Ray Registration of Vertebral Bodies for Image-Guided Spine Surgery Spine. ,vol. 29, pp. 293- 297 ,(2004) , 10.1097/01.BRS.0000106980.33936.AD
V. Rasche, B. Schreiber, C. Graeff, T. Istel, H. Schomberg, M. Grass, R. Koppe, E. Klotz, G. Rose, Performance of image intensifier-equipped X-ray systems for three-dimensional imaging computer assisted radiology and surgery. ,vol. 1256, pp. 187- 192 ,(2003) , 10.1016/S0531-5131(03)00321-2
D. Tomazevic, B. Likar, T. Slivnik, F. Pernus, 3-D/2-D registration of CT and MR to X-ray images IEEE Transactions on Medical Imaging. ,vol. 22, pp. 1407- 1416 ,(2003) , 10.1109/TMI.2003.819277
P. Haaker, E. Klotz, R. Koppe, R. Linde, Real-time distortion correction of digital X-ray II/TV-systems: an application example for Digital Flashing Tomosynthesis (DFTS) International Journal of Cardiac Imaging. ,vol. 6, pp. 39- 45 ,(1990) , 10.1007/BF01798431
M Grass, R Koppe, E Klotz, R Proksa, M.H Kuhn, H Aerts, J Op de Beek, R Kemkers, Three-dimensional reconstruction of high contrast objects using C-arm image intensifier projection data. Computerized Medical Imaging and Graphics. ,vol. 23, pp. 311- 321 ,(1999) , 10.1016/S0895-6111(99)00028-2
J.M. Fitzpatrick, J.B. West, The distribution of target registration error in rigid-body point-based registration IEEE Transactions on Medical Imaging. ,vol. 20, pp. 917- 927 ,(2001) , 10.1109/42.952729
Graeme P. Penney, Jürgen Weese, John A. Little, Paul Desmedt, Derek L. G. Hill, David J. Hawkes, A comparison of similarity measures for use in 2-D-3-D medical image registration IEEE Transactions on Medical Imaging. ,vol. 17, pp. 586- 595 ,(1998) , 10.1109/42.730403
Jay West, J. Michael Fitzpatrick, Matthew Y. Wang, Benoit M. Dawant, Calvin R. Maurer, Robert M. Kessler, Robert J. Maciunas, Christian Barillot, Didier Lemoine, André Collignon, Frederik Maes, Paul Suetens, Dirk Vandermeulen, Petra A. van den Elsen, Sandy Napel, Thilaka S. Sumanaweera, Beth Harkness, Paul F. Hemler, Derek L. G. Hill, David J. Hawkes, Colin Studholme, J. B. Antoine Maintz, Max A. Viergever, Gregoire Malandain, Xavier Pennec, Marilyn E. Noz, Gerald Q. Maguire, Michael Pollack, Charles A. Pelizzari, Richard A. Robb, Dennis Hanson, Roger P. Woods, Comparison and Evaluation of Retrospective Intermodality Brain Image Registration Techniques Journal of Computer Assisted Tomography. ,vol. 21, pp. 554- 566 ,(1997) , 10.1097/00004728-199707000-00007