Use of a Laser-equipped Centrifugal Analyzer for Kinetic Measurement of Serum IgG

作者: Gregory J Buffone , John Savory , R E Cross

DOI: 10.1093/CLINCHEM/20.10.1320

关键词:

摘要: Use of a visible laser in an Aminco "Rotochem" centrifugal analyzer enables light-scattering measurements to be made with this instrument. The modification is technically simple, inexpensive, and requires no fundamental changes basic instrument design. laser-modified has been applied the measurement IgG serum. A two-point kinetic analysis used quantitate from standard curve: protein concentration vs. change intensity scattered light during fixed time interval. Fourteen 20-µl samples can simultaneously determined 1.9 min. No sample blank corrections are required. Within-run precision studies, based each case on 56 replicate measurements, yielded coefficients variation 3.8% 4.8% for normal abnormal pools, respectively.

参考文章(1)
H.E. Schultze, G. Schwick, Quantitative immunologische bestimmung von plasmaproteinen Clinica Chimica Acta. ,vol. 4, pp. 15- 25 ,(1959) , 10.1016/0009-8981(59)90075-0