Specification and calibration of a multi-frequency MEIT system for stroke

作者: RJ Yerworth , RH Bayford , A McEwan , DS Holder , L Horesh

DOI:

关键词:

摘要: The performance of a modified MEIT system was investigated, using resistor phantom, for the imaging stroke based on specifications revealed in recent modelling (Horesh et al, 2005). Using ideal calibration does not meet as variation over load is 0.9%, frequency +/-1.5% and electrode combinations +/-2% exceeding better than 1% requirement 20-500kHz 1 to 66 Ω. However this has been useful gastric emptying (Soulsby 2005); extending lower frequencies should aid application along with further epilepsy trials (Fabrizi Introduction Time difference available there usually no opportunity take reference image. Frequency one method that may be suitable; however, relies spectrally consistent EIT measurements different electrodes. Unfortunately, requirements image appear demanding detailed anatomical mesh head data from literature, they estimated an accuracy less range 10 Hz – 500 kHz <1-70 Ω required. In our group, we have developing system, module Sheffield Mark 3.5 (Wilson 2000), but cross point switch permit addressing up 64 electrodes (Yerworth 2003). However, human recordings encountered unexpected difficulties (Romsauerova al purpose work revisit ascertain if it met recently by precise modelling. Characterization achieved measurement phantom; evaluated loads frequencies.

参考文章(5)
A J Wilson, P Milnes, A R Waterworth, R H Smallwood, B H Brown, Mk3.5: a modular, multi-frequency successor to the Mk3a EIS/EIT system. Physiological Measurement. ,vol. 22, pp. 49- 54 ,(2001) , 10.1088/0967-3334/22/1/307
R J Yerworth, R H Bayford, B Brown, P Milnes, M Conway, D S Holder, Electrical impedance tomography spectroscopy (EITS) for human head imaging Physiological Measurement. ,vol. 24, pp. 477- 489 ,(2003) , 10.1088/0967-3334/24/2/358
Tatsuma Yamamoto, Yoshitake Yamamoto, Electrical properties of the epidermal stratum corneum Medical & Biological Engineering & Computing. ,vol. 14, pp. 151- 158 ,(1976) , 10.1007/BF02478741
S Arridge, A Romsauerova, O Gilad, D Holder, L Horesh, Stroke type detection by multi-frequency electrical impedance tomography MFEIT—a feasibility study Proc. 6th Conf. on Biomedical Applications of Electrical Impedance Tomography (University College London) (2005). ,(2005)