Turbulent stress measurements downstream of six mechanical aortic valves in a pulsatile flow model.

作者: J.M. Hasenkam , H. Nygaard , M. Giersiepen , H. Reul , H. Stødkilde-Jørgensen

DOI: 10.1016/0021-9290(88)90201-1

关键词:

摘要: Abstract In a pulsatile flow model aortic Bjork-Shiley Standard, Convex-Concave and Monostrut valves were investigated together with the Hall-Kaster (Medtronic-Hall), St Jude Medical Starr-Edwards Silastic Ball valve using hot-film anemometry. Three-dimensional visualization of average systolic Reynolds normal stresses (RNS) reflected design valves. Mean RNS used for comparison fluid dynamic performance along Velocity Energy Ratio (VER 100 ) Turbulence (TER) as relative turbulence intensity flow. ranged from 13.2 to 37.6 Nm −2 all highest levels Standard lowest values in between.

参考文章(29)
R. S. Figliola, T. J. Mueller, On the Hemolytic and Thrombogenic Potential of Occluder Prosthetic Heart Valves From In-Vitro Measurements Journal of Biomechanical Engineering-transactions of The Asme. ,vol. 103, pp. 83- 90 ,(1981) , 10.1115/1.3138265
H. Reul, N. Talukder, E.W. Mu¨ller, Fluid mechanics of the natural mitral valve Journal of Biomechanics. ,vol. 14, pp. 361- 372 ,(1981) , 10.1016/0021-9290(81)90046-4
D.P. Giddens, R.I. Kitney, Extraction and characterisation of underlying velocity waveforms in poststenotic flow IEE Proceedings A Physical Science, Measurement and Instrumentation, Management and Education, Reviews. ,vol. 129, pp. 651- 662 ,(1982) , 10.1049/IP-A-1:19820103
Withaya Yongchareon, Donald F. Young, Initiation of turbulence in models of arterial stenoses Journal of Biomechanics. ,vol. 12, pp. 185- 196 ,(1979) , 10.1016/0021-9290(79)90141-6
Ajit P. Yoganathan, William H. Corcoran, Earl C. Harrison, In vitro velocity measurements in the vicinity of aortic prostheses Journal of Biomechanics. ,vol. 12, pp. 135- 152 ,(1979) , 10.1016/0021-9290(79)90153-2
W B Tindale, M M Black, T R P Martin, In vitro evaluation of prosthetic heart valves: anomalies and limitations. Clinical Physics and Physiological Measurement. ,vol. 3, pp. 115- 130 ,(1982) , 10.1088/0143-0815/3/2/002
Peter Kildeberg Paulsen, The Hot-Film Anemometer – a Method for Blood Velocity Determination European Surgical Research. ,vol. 12, pp. 140- 148 ,(1980) , 10.1159/000128119
J.Michael Hasenkam, Dieter Westphal, Helmut Reul, Jorgen Gormsen, Martin Giersiepen, Hans Stodkilde-Jorgensen, Peter Kildeberg Paulsen, Three-dimensional visualization of axial velocity profiles downstream of six different mechanical aortic valve prostheses, measured with a hot-film anemometer in a steady state flow model Journal of Biomechanics. ,vol. 20, pp. 353- 364 ,(1987) , 10.1016/0021-9290(87)90043-1
Frederick J. Walburn, Hani N. Sabbah, Paul D. Stein, Estimation of reynolds shear stresses during pulsatile flow in the region of aortic valves Annals of Biomedical Engineering. ,vol. 13, pp. 17- 23 ,(1985) , 10.1007/BF02371247