作者: Marshall M. Siegel , Keiko Tabei , Frank Lambert , Leonard Candela , Bart Zoltan
DOI: 10.1016/S1044-0305(98)00085-3
关键词:
摘要: The atmospheric pressure ionization (API) source for a commercial mass spectrometer was modified to operate as dual in both the electrospray (ESI) and chemical (APCI) techniques by simultaneously utilizing probe corona discharge needle. A switching box designed either manual or programmable modes permit rapid between without changing sources, probes, breaking vacuum. can be operated using following techniques: ESI only, APCI ESI/APCI simultaneously, alternatingly. optimum operating conditions these were similar manufacturer’s original specifications except that flow rate lower (∼50 µL/min versus 1000 µL/min) externally heated nebulizing gas found desirable. four-component mixture, introduced injection, used demonstrate versatility of source.