Fidelity of micromanometer pressure transduction in closed-end angiographic catheters

作者: Thomas R. Lloyd , Richard L. Donnerstein

DOI: 10.1016/0002-9149(89)90885-0

关键词:

摘要: Abstract Clinical high fidelity pressure recording has been greatly facilitated by the recent introduction of 2Fr micromanometer-tipped catheters. They can potentially be placed in any structure reached an open-end catheter ≥0.03-inch internal diameter. Because catheters often are not routinely necessary pediatric heart catheterization, we sought to determine whether accurate recordings could obtained from inserted into lumina closed-end angiographic

参考文章(2)
Gail G. Shapiro, L.Jerome Krovetz, Damped and undamped frequency responses of underdamped catheter manometer systems. American Heart Journal. ,vol. 80, pp. 226- 236 ,(1970) , 10.1016/0002-8703(70)90172-9
S. A. Glantz, J. V. Tyberg, Determination of frequency response from step response: application to fluid-filled catheters. American Journal of Physiology-heart and Circulatory Physiology. ,vol. 236, ,(1979) , 10.1152/AJPHEART.1979.236.2.H376