Damped and undamped frequency responses of underdamped catheter manometer systems.

作者: Gail G. Shapiro , L.Jerome Krovetz

DOI: 10.1016/0002-8703(70)90172-9

关键词:

摘要: Abstract Compliance is the most troublesome variable in obtaining high frequency response. Factors such as occult air bubbles and leaky connections cause greatest difficulty. Damped natural was shown to be directly related radius inversely square root of length catheter. Coils, bends, angle attachment catheter transducer are minor importance a good