The radiotherapy clinical trials projects at the ESRF: Technical aspects

作者: M. Renier , Th. Brochard , C. Nemoz , H. Requardt , E. Bräuer

DOI: 10.1016/J.EJRAD.2008.04.057

关键词:

摘要: The radiotherapy clinical trials projects, both aiming at treating aggressive brain tumors, require several major modifications and new constructions the ESRF ID17 Biomedical beamline. application of Stereotactic Synchrotron Radiation Therapy (SSRT) technique mainly necessitates an upgrade existing patient positioning system, which was formerly used for angiography program. It will allow accurate positioning, translation rotation during treatment. For Microbeam (MRT) project, a white beam hutch be constructed to accommodate dedicated system. Consequently, control hutches related installations also completely refurbished. Furthermore, foreseen installation second X-ray source, doubling currently available photon flux high energies, requires redesign most optical components handle increased power densities. Starting from current beamline layout, paper present update different modification/construction including general organization planning.

参考文章(11)
Jean-François Adam, Aurélie Joubert, Marie-Claude Biston, Anne-Marie Charvet, Michel Peoc’h, Jean-François Le Bas, Jacques Balosso, François Estève, Hélène Elleaume, Prolonged survival of Fischer rats bearing F98 glioma after iodine-enhanced synchrotron stereotactic radiotherapy. International Journal of Radiation Oncology Biology Physics. ,vol. 64, pp. 603- 611 ,(2006) , 10.1016/J.IJROBP.2005.09.004
E Bräuer-Krisch, A Bravin, L Zhang, E Siegbahn, J Stepanek, H Blattmann, DN Slatkin, J-O Gebbers, M Jasmin, JA Laissue, Tecomet Subsidiary, Erratum: “Characterization of a tungsten/gas multislit collimator (TMSC) for microbeam radiation therapy at the European Synchrotron Radiation Facility”[Rev. Sci. Instrum. 76, 064303 (2005)] Review of Scientific Instruments. ,vol. 77, pp. 039901- ,(2006) , 10.1063/1.2173954
M Miura, H Blattmann, E Bräuer-Krisch, A Bravin, A L Hanson, M M Nawrocky, P L Micca, D N Slatkin, J A Laissue, Radiosurgical palliation of aggressive murine SCCVII squamous cell carcinomas using synchrotron-generated X-ray microbeams British Journal of Radiology. ,vol. 79, pp. 71- 75 ,(2006) , 10.1259/BJR/50464795
Y. Dabin, A. Draperi, H. Elleaume, A-M. Charvet, T. Brochard, M. Perez, C. Nemoz, G. Blattmann, M. Renier, F. Fournier, J-L Dupuy, B. Lemoine, P. Bouhaniche, W. Thomlinson, P. Suortti, A patient positioning system for the ESRF medical imaging facility Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment. ,vol. 467, pp. 1342- 1345 ,(2001) , 10.1016/S0168-9002(01)00657-X
H. Elleaume, A.M. Charvet, P. Berkvens, G. Berruyer, T. Brochard, Y. Dabin, M.C. Dominguez, A. Draperi, S. Fiedler, G. Goujon, G. Le Duc, M. Mattenet, C. Nemoz, M. Perez, M. Renier, C. Schulze, P. Spanne, P. Suortti, W. Thomlinson, F. Esteve, B. Bertrand, J.F. Le Bas, Instrumentation of the ESRF medical imaging facility Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment. ,vol. 428, pp. 513- 527 ,(1999) , 10.1016/S0168-9002(99)00167-9
P. Suortti, S. Fiedler, A. Bravin, T. Brochard, M. Mattenet, M. Renier, P. Spanne, W. Thomlinson, A. M. Charvet, H. Elleaume, C. Schulze-Briese, A. C. Thompson, Fixed-exit monochromator for computed tomography with synchrotron radiation at energies 18–90 keV Journal of Synchrotron Radiation. ,vol. 7, pp. 340- 347 ,(2000) , 10.1107/S0909049500008384
J. Braun, H. Bruhns, M. Trinczek, J. R. Crespo López-Urrutia, J. Ullrich, Novel technique for high-precision Bragg-angle determination in crystal x-ray spectroscopy Review of Scientific Instruments. ,vol. 76, pp. 073105- 6 ,(2005) , 10.1063/1.1947878
M Renier, T Brochard, C Nemoz, W Thomlinson, A white-beam fast-shutter for microbeam radiation therapy at the ESRF Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment. ,vol. 479, pp. 656- 660 ,(2002) , 10.1016/S0168-9002(01)00905-6
Marie-Claude Biston, Aurélie Joubert, Jean-François Adam, Hélène Elleaume, Sylvain Bohic, Anne-Marie Charvet, François Estève, Nicolas Foray, Jacques Balosso, Cure of Fisher Rats Bearing Radioresistant F98 Glioma Treated with cis-Platinum and Irradiated with Monochromatic Synchrotron X-Rays Cancer Research. ,vol. 64, pp. 2317- 2323 ,(2004) , 10.1158/0008-5472.CAN-03-3600
Paola Coan, Angela Peterzol, Stefan Fiedler, Cyril Ponchut, Jean Claude Labiche, Alberto Bravin, Evaluation of imaging performance of a taper optics CCD; FReLoN' camera designed for medical imaging. Journal of Synchrotron Radiation. ,vol. 13, pp. 260- 270 ,(2006) , 10.1107/S0909049506008983