Electron fluence correction factors for conversion of dose in plastic to dose in water.

作者: G. X. Ding , D. W. O. Rogers , J. E. Cygler , T. R. Mackie

DOI: 10.1118/1.597930

关键词:

摘要: In radiationdosimetry protocols, plastic is allowed as a phantom material for the determination of absorbed dose to water in electron beams. The fluence correction factor needed conversion measured water. There are large discrepancies among recommended values well factors when polystyrene used material. Using Monte Carlo technique, we have calculated incident clinical beam energies between 5 and 50 MeV function depth clear polystyrene, white PMMA materials compared results with those protocols experimental from published data. calculations, beams simulated using EGS4 user-code BEAM variety medical accelerators. study shows that our factor, φ p w , energy Ē o little dependence on other aspects quality. However at d max indirectly influenced by quality since they vary also varies Calculated range 1.005–1.045 phantom, 1.005–1.038 0.996–1.016 phantom. Our about 1–2% higher than determined according AAPM TG-25 protocol or polystyrene. explain some variations data because its dependence. A simple formula derived which gives R 0.6 −0.1 any 25 three plastics: PMMA. makes careful distinction corresponding IAEA Code Practice quantity, h m .

参考文章(20)
I A D Bruinvis, S Heukelom, B J Mijnheer, Comparison of ionisation measurements in water and polystyrene for electron beam dosimetry Physics in Medicine and Biology. ,vol. 30, pp. 1043- 1053 ,(1985) , 10.1088/0031-9155/30/10/003
Alex F. Bielajew, D.W.O. Rogers, Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transport Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. ,vol. 18, pp. 165- 181 ,(1986) , 10.1016/S0168-583X(86)80027-1
George X. Ding, Shlomo Shalev, Georgi Gluchev, A ρ-θ technique for treatment verification in radiotherapy and its clinical applications Medical Physics. ,vol. 20, pp. 1135- 1143 ,(1993) , 10.1118/1.596967
S C Klevenhagen, Implication of electron backscattering for electron dosimetry Physics in Medicine and Biology. ,vol. 36, pp. 1013- 1018 ,(1991) , 10.1088/0031-9155/36/7/009
Kenneth R. Kase, Gary J. Adler, Bengt E. Bjärngard, Comparisons of electron beam dose measurements in water and polystyrene using various dosimeters Medical Physics. ,vol. 9, pp. 13- 19 ,(1982) , 10.1118/1.595055
J. Alan Rawlinson, Alex F. Bielajew, Peter Munro, Duncan M. Galbraith, Theoretical and experimental investigation of dose enhancement due to charge storage in electron-irradiated phantoms Medical Physics. ,vol. 11, pp. 814- 821 ,(1984) , 10.1118/1.595569
G. X. Ding, D. W. O. Rogers, T. R. Mackie, Mean energy, energy-range relationships and depth-scaling factors for clinical electron beams. Medical Physics. ,vol. 23, pp. 361- 376 ,(1996) , 10.1118/1.597788
L. O. Mattsson, Alan E. Nahum, Concerning comparisons of electron beam absorbed dose measurements in water and polystyrene Medical Physics. ,vol. 11, pp. 85- 86 ,(1984) , 10.1118/1.595464