作者: Gaurav Bandi , Charles C. Wen , Eric A. Wilkinson , Sean P. Hedican , Timothy D. Moon
关键词:
摘要: Purpose: To compare tip temperatures of the vibrating blades newer-generation Harmonic Ace device and older-generation Ultracision Scalpel LCS-K5. Materials Methods: Comparison two different harmonic scalpel was performed by applying them to 3-mm strips raw chicken breast at room temperature measuring blade 0.5-second intervals using a thermocouple. Following activation, allowed passively cool 30°C each time before reactivation. Each tested power settings 3 5 after activation for 1 seconds. All measurements were times. Results: Compared LCS-K5, associated with higher level all times (1–5 seconds), longer (3–5 seconds). The also takes reach peak can...