Comparison of blade temperature dynamics after activation of Harmonic Ace scalpel and the Ultracision Harmonic Scalpel LCS-K5.

作者: Gaurav Bandi , Charles C. Wen , Eric A. Wilkinson , Sean P. Hedican , Timothy D. Moon

DOI: 10.1089/END.2007.0169

关键词:

摘要: Purpose: To compare tip temperatures of the vibrating blades newer-generation Harmonic Ace device and older-generation Ultracision Scalpel LCS-K5. Materials Methods: Comparison two different harmonic scalpel was performed by applying them to 3-mm strips raw chicken breast at room temperature measuring blade 0.5-second intervals using a thermocouple. Following activation, allowed passively cool 30°C each time before reactivation. Each tested power settings 3 5 after activation for 1 seconds. All measurements were times. Results: Compared LCS-K5, associated with higher level all times (1–5 seconds), longer (3–5 seconds). The also takes reach peak can...

参考文章(8)
K.M Kadesky, B Schopf, J.F Magee, G.K Blair, Proximity injury by the ultrasonically activated scalpel during dissection. Journal of Pediatric Surgery. ,vol. 32, pp. 878- 879 ,(1997) , 10.1016/S0022-3468(97)90641-2
J AWWAD, K ISAACSON, The harmonic scalpel: An intraoperative complication Obstetrics & Gynecology. ,vol. 88, pp. 718- 720 ,(1996) , 10.1016/0029-7844(96)00075-0
R Antonutti, I Fontes-Dislaire, J.L Rumeau, D Mutter, G Fourtanier, Étude expérimentale comparant bistouri électrique monopolaire et bistouri ultrasonique Annales De Chirurgie. ,vol. 126, pp. 330- 335 ,(2001) , 10.1016/S0003-3944(01)00515-6
T. Owaki, S. Nakano, K. Arimura, T. Aikou, The ultrasonic coagulating and cutting system injures nerve function. Endoscopy. ,vol. 34, pp. 575- 579 ,(2002) , 10.1055/S-2002-33221
J. Carlander, K. Johansson, S. Lindström, Å. K. Velin, C. H. Jiang, C. Nordborg, Comparison of experimental nerve injury caused by ultrasonically activated scalpel and electrosurgery British Journal of Surgery. ,vol. 92, pp. 772- 777 ,(2005) , 10.1002/BJS.4948
T. Kinoshita, E. Kanehira, K. Omura, K. Kawakami, Y. Watanabe, Experimental study on heat production by a 23.5-kHz ultrasonically activated device for endoscopic surgery. Surgical Endoscopy and Other Interventional Techniques. ,vol. 13, pp. 621- 625 ,(1999) , 10.1007/S004649901055
Dongsheng Xu, Martin Pollock, Experimental nerve thermal injury Brain. ,vol. 117, pp. 375- 384 ,(1994) , 10.1093/BRAIN/117.2.375
U Cagirici, M Cikirikcioglu, H Posacioglu, A Cakan, A Veral, The histological effect of harmonic scalpel and electrocautery in lung resections. An experimental study in a rat model. Journal of Cardiovascular Surgery. ,vol. 45, pp. 63- 65 ,(2004)