Mercury Sphygmomanometers Should Not be Abandoned: An Advisory Statement From the Council for High Blood Pressure Research, American Heart Association

作者: Daniel W. Jones , Edward D. Frohlich , Carlene M. Grim , Clarence E. Grim , Kathryn A. Taubert

DOI: 10.1161/01.HYP.37.2.185

关键词:

摘要: In healthcare institutions around this country and the world, mercury sphygmomanometers are being removed.1 many situations, decision to replace instruments is made without significant input from involved clinicians or consideration of health risks that will follow if they replaced by less accurate devices. Blood pressure measurement an important indicator current clinical condition patients a powerful predictor future cardiovascular overall health.2 Blood often considered “routine” performed those with least training. institutions, blood low priority, than ideal quality control related equipment selection, calibration repair, personnel training performance. For more century, gravity sphygmomanometer has been gold standard for indirect pressure. Indeed, world primary manometer. It simple, gravity-based unit easy calibration, infrequent need it validated in circumstances against direct intra-arterial measurement.3 In recent years, these units have aneroid recently electronic manometers. Justifications replacement manometers included concerns about safety mercury, regulations regarding use …

参考文章(8)
A C Rennie, M McGregor-Schuerman, I M Dale, C Robinson, R McWilliam, Mercury poisoning after spillage at home from a sphygmomanometer on loan from hospital. BMJ. ,vol. 319, pp. 366- 367 ,(1999) , 10.1136/BMJ.319.7206.366
ND Markandu, F Whitcher, A Arnold, C Carney, The mercury sphygmomanometer should be abandoned before it is proscribed. Journal of Human Hypertension. ,vol. 14, pp. 31- 36 ,(2000) , 10.1038/SJ.JHH.1000932
Sl Carney, Ahb Gillies, Sl Green, O Paterson, Ms Taylor, Aj Smith, Hospital blood pressure measurement: staff and device assessment. Journal of Quality in Clinical Practice. ,vol. 19, pp. 95- 98 ,(1999) , 10.1046/J.1440-1762.1999.00308.X
S. A. Yarows, R. D. Brook, Measurement variation among 12 electronic home blood pressure monitors American Journal of Hypertension. ,vol. 13, pp. 276- 282 ,(2000) , 10.1016/S0895-7061(99)00182-X
NJ Langford, RE Ferner, Toxicity of mercury. Journal of Human Hypertension. ,vol. 13, pp. 651- 656 ,(1999) , 10.1038/SJ.JHH.1000896
D Perloff, C Grim, J Flack, E D Frohlich, M Hill, M McDonald, B Z Morgenstern, Human blood pressure determination by sphygmomanometry. Circulation. ,vol. 88, pp. 2460- 2470 ,(1993) , 10.1161/01.CIR.88.5.2460
O'Brien E, Will mercury manometers soon be obsolete Journal of Human Hypertension. ,vol. 9, pp. 933- 934 ,(1995)