Paper spray and extraction spray mass spectrometry for the direct and simultaneous quantification of eight drugs of abuse in whole blood.

作者: Ryan D. Espy , Sebastiaan Frans Teunissen , Nicholas E. Manicke , Yue Ren , Zheng Ouyang

DOI: 10.1021/AC5016408

关键词:

摘要: Determination of eight drugs abuse in blood has been performed using paper spray or extraction mass spectrometry under 2 min with minimal sample preparation. A method optimized for quantification amphetamine, methamphetamine, 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxy-N-methylamphetamine (MDMA), 3,4-methylenedioxy-N-ethylamphetamine (MDEA), morphine, cocaine, and Δ9-tetrahydrocannabinol (THC) from a single spot. Sample to variations 1–5% relative standard deviation were achieved stable isotope-labeled internal standards tandem spectrometry. Limits detection all below typical physiological toxicological levels. Paper each used less than 10 μL whole blood. These methods exhibit the potential performing rapid high-throughput assays selective on-site multicompound quantitative screening illicit drugs.

参考文章(32)
Le Long, Joseph Kost, Samir Mitragotri, Combined Effect of Low-Frequency Ultrasound and Iontophoresis: Applications for Transdermal Heparin Delivery Pharmaceutical Research. ,vol. 17, pp. 1151- 1154 ,(2000) , 10.1023/A:1026426317413
Joanne E. Brady, Guohua Li, Prevalence of Alcohol and Other Drugs in Fatally Injured Drivers Addiction. ,vol. 108, pp. 104- 114 ,(2013) , 10.1111/J.1360-0443.2012.03993.X
Thomas Kraemer, Liane D. Paul, Bioanalytical procedures for determination of drugs of abuse in blood Analytical and Bioanalytical Chemistry. ,vol. 388, pp. 1415- 1435 ,(2007) , 10.1007/S00216-007-1271-6
Frank T. Peters, Olaf H. Drummer, Frank Musshoff, Validation of new methods. Forensic Science International. ,vol. 165, pp. 216- 224 ,(2007) , 10.1016/J.FORSCIINT.2006.05.021
Zhiping Zhang, Wei Xu, Nicholas E. Manicke, R. Graham Cooks, Zheng Ouyang, Silica Coated Paper Substrate for Paper-Spray Analysis of Therapeutic Drugs in Dried Blood Spots Analytical Chemistry. ,vol. 84, pp. 931- 938 ,(2012) , 10.1021/AC202058W
Nicholas Edward Manicke, Paul Abu-Rabie, Neil Spooner, Zheng Ouyang, R Graham Cooks, None, Quantitative analysis of therapeutic drugs in dried blood spot samples by paper spray mass spectrometry: an avenue to therapeutic drug monitoring. Journal of the American Society for Mass Spectrometry. ,vol. 22, pp. 1501- 1507 ,(2011) , 10.1007/S13361-011-0177-X
He Wang, Yue Ren, Morgan N. McLuckey, Nicholas E. Manicke, Jonghyuck Park, Lingxing Zheng, Riyi Shi, R. Graham Cooks, Zheng Ouyang, Direct Quantitative Analysis of Nicotine Alkaloids from Biofluid Samples using Paper Spray Mass Spectrometry Analytical Chemistry. ,vol. 85, pp. 11540- 11544 ,(2013) , 10.1021/AC402798M
S. Steinmeyer, H. Ohr, H.J. Maurer, M.R. Moeller, Practical aspects of roadside tests for administrative traffic offences in Germany. Forensic Science International. ,vol. 121, pp. 33- 36 ,(2001) , 10.1016/S0379-0738(01)00450-9
Demian R. Ifa, Ayanna U. Jackson, Giuseppe Paglia, R. Graham Cooks, Forensic applications of ambient ionization mass spectrometry. Analytical and Bioanalytical Chemistry. ,vol. 394, pp. 1995- 2008 ,(2009) , 10.1007/S00216-009-2659-2