Wedge-shaped scintillation crystals for positron emission tomography.

作者: R. Lecomte , D. Schmitt , E. LeBel

DOI:

关键词:

摘要: The use of wedge-shaped scintillation crystals for positron emission tomography is reevaluated with the aim delimiting its range benefit. A linear attenuation simulation model used to generate detector geometric aperture functions, and investigation carried out in terms a newly introduced shape-independent definition spatial resolution, "spectral resolution." Contrary previous expectations, it concluded that shaped do not improve performance high resolution detection systems, either or without intercrystal septa. However, wedges are found be useful lower systems boundaries applicability estimated NaI, BaF2, GSO, BGO scintillators.

参考文章(6)
S.E. Derenzo, T.F. Budinger, R.H. Huesman, T. Vuletich, J.L. Cahoon, High resolution positron emission tomography using 3 mm wide bismuth germanate crystals The Journal of Nuclear Medicine. ,(1984)
Timothy J. Holmes, David C. Ficke, Analysis of Positron-Emission Tomography Scintillation-Detectors with Wedge Faces and Inter-Crystal Septa IEEE Transactions on Nuclear Science. ,vol. 32, pp. 826- 830 ,(1985) , 10.1109/TNS.1985.4336949
Robert F. Wagner, Toward a unified view of radiological imaging systems. Part II: Noisy images Medical Physics. ,vol. 4, pp. 279- 296 ,(1974) , 10.1118/1.594362
R. Lecomte, D. Schmitt, G. Lamoureux, Geometry Study of a High Resolution PET Detection System Using Small Detectors IEEE Transactions on Nuclear Science. ,vol. 31, pp. 556- 561 ,(1984) , 10.1109/TNS.1984.4333318
J. M. Roney, C. J. Thompson, Detector Identification with Four BGO Crystals on a Dual PMT IEEE Transactions on Nuclear Science. ,vol. 31, pp. 1022- 1027 ,(1984) , 10.1109/TNS.1984.4333434
C. Nahmias, D. B. Kenyon, E. S. Garnett, Experience with a High Efficiency Positron Emission Tomograph IEEE Transactions on Nuclear Science. ,vol. 29, pp. 548- 550 ,(1982) , 10.1109/TNS.1982.4335905