An improved analytical detector response function model for multilayer small-diameter PET scanners

作者: D Strul , R B Slates , M Dahlbom , S R Cherry , P K Marsden

DOI: 10.1088/0031-9155/48/8/302

关键词:

摘要: The optimization of spatial resolution is a critical consideration in the design small-diameter positron emission tomography (PET) scanners for animal imaging, and often addressed with Monte Carlo simulations. As faster simpler solution, we have developed new analytical model PET detector response function, implemented small single-slice, multilayer scanner. accuracy has been assessed by comparison both simulations experimental measurements published literature. Results from agreed well method, being noise free two to three orders magnitude faster. only major discrepancy was slight underestimation width point spread function method as inter-crystal scatter neglected. We observed good agreement between predictions measurements. For large-diameter additional discrepancies were seen due photon acollinearity, which not considered model. shown that simple fast can provide accurate estimates scanners, could be useful tool several applications, complementing or cross-validating other simulation methods.

参考文章(31)
P. D. Cutler, Simon R Cherry, W. M. Digby, E. J. Hoffman, M. E. Phelps, Design features and performance of a PET system for animal research. The Journal of Nuclear Medicine. ,vol. 33, pp. 595- 604 ,(1992)
Arion F. Chatziioannou, Thomas H. Farquhar, Simon R. Cherry, Robert W. Silverman, Michael E. Phelps, Yiping Shao, Ken Meadors, Marjan Pedarsani, Performance evaluation of microPET: A high-resolution lutetium oxyorthosilicate PET scanner for animal imaging The Journal of Nuclear Medicine. ,vol. 40, pp. 1164- 1175 ,(1999)
Olivier G. Rousset, Yilong Ma, Alan C. Evans, Correction for Partial Volume Effects in PET: Principle and Validation The Journal of Nuclear Medicine. ,vol. 39, pp. 904- 911 ,(1998)
Pamela B. Garlick, Paul K. Marsden, Alison C. Cave, Harold G. Parkes, Randal Slates, Yiping Shao, Robert W. Silverman, Simon R. Cherry, PET and NMR dual acquisition (PANDA): applications to isolated, perfused rat hearts. NMR in Biomedicine. ,vol. 10, pp. 138- 142 ,(1997) , 10.1002/(SICI)1099-1492(199705)10:3<138::AID-NBM474>3.0.CO;2-H
Timothy J. Holmes, David C. Ficke, Analysis of Positron-Emission Tomography Scintillation-Detectors with Wedge Faces and Inter-Crystal Septa IEEE Transactions on Nuclear Science. ,vol. 32, pp. 826- 830 ,(1985) , 10.1109/TNS.1985.4336949
Neil A. Keller, Lawrence R. Lupton, PET Detector Ring Aperture Function Calculations Using Monte Carlo Techniques IEEE Transactions on Nuclear Science. ,vol. 30, pp. 676- 680 ,(1983) , 10.1109/TNS.1983.4332353
Gunnar Brix, Josef Doll, Matthias E. Bellemann, Herbert Trojan, Uwe Haberkorn, Peter Schmidlin, Hermann Ostertag, Use of scanner characteristics in iterative image reconstruction for high-resolution positron emission tomography studies of small animals European Journal of Nuclear Medicine and Molecular Imaging. ,vol. 24, pp. 779- 786 ,(1997) , 10.1007/S002590050120
S. E. Derenzo, R. H. Huesman, J. L. Cahoon, A. Geyer, D. Uber, T. Vuletich, T. F. Budinger, Initial Results from the Donner 600 Crystal Positron Tomograph IEEE Transactions on Nuclear Science. ,vol. 34, pp. 321- 325 ,(1987) , 10.1109/TNS.1987.4337356
R. Lecomte, D. Schmitt, G. Lamoureux, Geometry Study of a High Resolution PET Detection System Using Small Detectors IEEE Transactions on Nuclear Science. ,vol. 31, pp. 556- 561 ,(1984) , 10.1109/TNS.1984.4333318
G. Tsang, C. Moisan, J.G. Rogers, A simulation to model position encoding multicrystal PET detectors IEEE Transactions on Nuclear Science. ,vol. 42, pp. 2236- 2243 ,(1995) , 10.1109/23.489420