New Interpolation Formula for Carbon Resistance Thermometry

作者: M. C. Hetzler , D. Walton

DOI: 10.1063/1.1683196

关键词:

摘要: A new interpolation formula for use in carbon resistance thermometry is suggested, and the results of fitting to experimental R vs T data resistors manufactured by Speer Company are described. The maximum difference between calculated measured temperatures was 1%. also found give good agreement with when extrapolated somewhat above or below region covered calibration points.

参考文章(3)
J. R. Clement, E. H. Quinnell, The Low Temperature Characteristics of Carbon‐Composition Thermometers Review of Scientific Instruments. ,vol. 23, pp. 213- 216 ,(1952) , 10.1063/1.1746229
A. Brown, M. W. Zemansky, H. A. Boorse, A Sensitive and Reproducible Thermometer in the Range 2° to 20°K Physical Review. ,vol. 84, pp. 1050- 1050 ,(1951) , 10.1103/PHYSREV.84.1050
O. V. Lounasmaa, CORRESPONDENCE A Simple Formula for use with Carbon Thermometers at Low Temperatures Philosophical Magazine. ,vol. 3, pp. 652- 660 ,(1958) , 10.1080/14786435808565806