Intraoperative decrease in pulse oximeter readings following injection of isosulfan blue

作者: Robert W. Hoskin , Robert Granger

DOI: 10.1007/BF03019812

关键词:

摘要: Purpose: Sentinel lymph node mapping, involving injection of isosulfan blue dye around a tumour, is beginning to be used in patients with carcinoma the breast. Absorption into circulation may interfere pulse oximetry, causing falsely low readings. This report describes changes oximeter readings following for sentinel mapping patient