作者: Rolf W. Sparidans , Jan M. Prins , Jan H.M. Schellens , Jos H. Beijnen
DOI: 10.1002/BMC.797
关键词:
摘要: A liquid chromatography-tandem mass spectrometric assay for the determination of antiretroviral nucleoside emtricitabine in human plasma was developed and validated using a simple sample pre-treatment procedure. After addition 5'-deoxy-5-fluorocytidine as internal standard protein precipitation with acetonitrile, supernatant directly injected isocratic chromatographic system polar embedded reversed-phase column formic acid water-methanol eluent. The eluate completely led into an electrospray interface positive ionization analytes were quantified triple quadrupole spectrometry. range 5-5000 ng/mL. Intra-day precisions