A novel device to evaluate the stiffness of ankle-foot orthosis devices.

作者: P. Cappa, , F. Patane`, , M. M. Pierro

DOI: 10.1115/1.1634993

关键词:

摘要: The evaluation of mechanical behavior plastic Ankle-Foot Orthosis (AFO) is important since AFO can provide an efficient support to patients with disabilities in locomotion. This paper reports on a novel testing apparatus that allows: (a) the stiffness sagittal and frontal planes; (b) conduction semi-automatic trials; and, finally, (c) global accuracy associated values always less than 4%. are determined by measurements imposed relative displacements between foot shank orthosis induced reaction forces. data collected together exact 2-D approach, those provided gait analysis systems, allows better understand alteration ankle orthosis, improve clinical management patients.

参考文章(27)
Narender P. Reddy, Tai-Ming Chu, Stress distribution in the ankle-foot orthosis used to correct pathological gait. Journal of Rehabilitation Research and Development. ,vol. 32, pp. 349- 360 ,(1995)
Dan Deitz, Optimizing orthotic designs with FEA Mechanical Engineering. ,vol. 119, pp. 70- 71 ,(1997)
D. Leone, S. Diemente, M. Lopez-Isa, Structural stability prediction for thermoplastic ankle-foot orthoses Proceedings of the 1991 IEEE Seventeenth Annual Northeast Bioengineering Conference. pp. 231- 232 ,(1991) , 10.1109/NEBC.1991.154659
David A. Winter, Aftab E. Patla, Francois Prince, Milad Ishac, Krystyna Gielo-Perczak, Stiffness Control of Balance in Quiet Standing Journal of Neurophysiology. ,vol. 80, pp. 1211- 1221 ,(1998) , 10.1152/JN.1998.80.3.1211
S Syngellakis, M A Arnold, H Rassoulian, Assessment of the non-linear behaviour of plastic ankle foot orthoses by the finite element method. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. ,vol. 214, pp. 527- 539 ,(2000) , 10.1243/0954411001535561
G. K. Cole, B. M. Nigg, J. L. Ronsky, M. R. Yeadon, Application of the joint coordinate system to three-dimensional joint attitude and movement representation: a standardization proposal. Journal of Biomechanical Engineering-transactions of The Asme. ,vol. 115, pp. 344- 349 ,(1993) , 10.1115/1.2895496
X.H. Duan, R.H. Allen, J.Q. Sun, A stiffness-varying model of human gait Medical Engineering & Physics. ,vol. 19, pp. 518- 524 ,(1997) , 10.1016/S1350-4533(97)00022-2
Susan Sienko Thomas, Cathleen E Buckon, Sabrina Jakobson-Huston, Michael D Sussman, Michael D Aiona, Stair locomotion in children with spastic hemiplegia: the impact of three different ankle foot orthosis (AFOS) configurations Gait & Posture. ,vol. 16, pp. 180- 187 ,(2002) , 10.1016/S0966-6362(02)00002-4
Thomas R. Lunsford, Thomas Ramm, Joseph A. Miller, Viscoelastic Properties of Plastic Pediatric AFOs Jpo Journal of Prosthetics and Orthotics. ,vol. 6, pp. 3- 9 ,(1994) , 10.1097/00008526-199400610-00002