A diffusion limited reaction theory for a microtiter plate assay.

作者: Manne Stenberg , Maria Werthén , Sten Theander , Håkan Nygren

DOI: 10.1016/0022-1759(88)90028-2

关键词:

摘要: Calculations are presented describing the diffusion limited kinetics of a solid-phase immunoassay in which reactants immobilized at inner surface cylindrical well. The calculations refer to an unstirred situation and simplified expressions can be used for calibration optimization assay.

参考文章(8)
Håkan Nygren, Maria Werthen, Manne Stenberg, Kinetics of antibody binding to solid-phase-immobilised antigen. Effect of diffusion rate limitation and steric interaction. Journal of Immunological Methods. ,vol. 101, pp. 63- 71 ,(1987) , 10.1016/0022-1759(87)90217-1
Horatio Scott Carslaw, John Conrad Jaeger, Conduction of Heat in Solids ,(1959)
Håkan Nygren, Manne Stenberg, Calibration by ellipsometry of the enzyme-linked immunosorbent assay. Journal of Immunological Methods. ,vol. 80, pp. 15- 24 ,(1985) , 10.1016/0022-1759(85)90159-0
Manne Stenberg, Lars Stiblert, Håkan Nygren, External diffusion in solid-phase immunoassays Journal of Theoretical Biology. ,vol. 120, pp. 129- 140 ,(1986) , 10.1016/S0022-5193(86)80169-2
Manne Stenberg, Håkan Nygren, A diffusion limited reaction theory for a solid-phase immunoassay Journal of Theoretical Biology. ,vol. 113, pp. 589- 597 ,(1985) , 10.1016/S0022-5193(85)80040-0
Manne Stenberg, Håkan Nygren, A receptor-ligand reaction studied by a novel analytical tool—The Isoscope ellipsometer Analytical Biochemistry. ,vol. 127, pp. 183- 192 ,(1982) , 10.1016/0003-2697(82)90163-4
Manne Stenberg, Hans Elwing, Håkan Nygren, Kinetics of reaction zone formation with radial diffusion of ligands over a receptor-coated surface. Journal of Theoretical Biology. ,vol. 98, pp. 307- 320 ,(1982) , 10.1016/0022-5193(82)90267-3