作者: A G Cherian , S J Soldin , J G Hill
DOI: 10.1093/CLINCHEM/27.5.748
关键词:
摘要: We automated the Jendrassik-Grof method for measurement of bilirubin with Greiner Selective Analyzer GSA II D. Comparison manual showed a highly significant correlation coefficient 0.994 total and 0.980 "direct" measurement. The good precision was linear to 667 mu mol/L, highest concentration tested. Addition hemolysate caused greater effect on direct than bilirubin. inhibitory bilirubin, though small, statistically at hemoglobin concentrations 2 g/L. involving diazotized 2,4-dichloroaniline be relatively less affected by hemolysis turbidity in