作者: Lejla Gurbeta , A. Badnjević
DOI: 10.1007/S12553-016-0154-2
关键词:
摘要: With increased sophistication of electrical medical devices and more dynamic working environment conditions, safety accuracy requirements are becoming strict. Healthcare institutions challenged in keeping their safe to use, accurate reliable terms measuring monitoring vital parameters. In healthcare maintaining overall operating functions at the required level performance can be achieved through periodical inspections. International guidelines, such as Directives IEC 60601, ISO 62353 MDD 93/42, define how should perform these checks. countries, where those guidelines not adopted, device inspections conducted accordance new approach. This paper presents software solution for tracking inspection process public private institutions. The is implemented Oracle Application Development Framework Technology (ADF) it used facilitate gathering documents Inspection Certificates, Working Orders, Measurement Reports, Calculated Errors, also keep track dates next inspection. accessed online via Laboratory website, all clients, well professional laboratory staff login using own username password which makes data confidential. validated Bosnia Herzegovina (BH). Out 331 health care BH, was 218 than 1800 tests reports were imported by date that this written.