作者: Hikaru Inoue , Yasukazu Yoshizawa , Tsuneo Morii
DOI: 10.1143/JPSJ.34.1437
关键词:
摘要: Gamma-ray energies lower than 500 keV and relative intensities have been precisely measured with Ge(Li) detectors. Many experimental effects such as the source position effect, counting rate room temperature etc. are examined. Results follows: 303.916±0.012 400.624±0.030 for 75 Se; 433.932±0.014 108 m Ag; 255.126±0.010 (3.33±0.13) 391.686±0.027 (100) 113 Sn; 364.456±0.031 131 I; 161 (0.98±0.07), 223 (0.76±0.05), 276 (11.6±0.5), 302.839±0.008 (29.6±1.1), 355.999±0.024 383.841±0.027 (14.9±0.6) 133 Ba. Parentheses show intensities.