A 28 μm water-vapor laser interferometer for plasma diagnostics

作者: S. Goto , S. Iwama , N. Satomi , M. Yamanaka , T. Ishimura

DOI: 10.1007/BF01009394

关键词:

摘要: A high-power CW 27.972 μm water-vapor laser has been developed for interferometric and polarimetric measurements of plasmas. The maximum output power the density in cavity with 2.0 cm bore 220 length reach 85 mW 109 μW/cm3, respectively. An interferometer using this is constructed its performance examined as a study on method magnetic field measurement. Preliminary experiments are also carried out compact toroidal plasma field-reversal configuration.

参考文章(3)
S. Iwama, N. Satomi, M. Yamanaka, S. Goto, T. Ishimura, H. Ito, A CW 28 μm water vapor laser for plasma diagnostics International Journal of Infrared and Millimeter Waves. ,vol. 2, pp. 1199- 1208 ,(1981) , 10.1007/BF01007084
I. S. Falconer, S. A. Ramsden, Measurement of the Faraday Rotation of a Laser Beam Passing through a θ-Pinch Plasma Journal of Applied Physics. ,vol. 39, pp. 3449- 3458 ,(1968) , 10.1063/1.1656796
M. Yamanaka, Y. Takeda, S. Tanigawa, A. Nishizawa, N. Noda, J. Fujita, M. Takai, M. Shimobayashi, Y. Hayashi, T. Koizumi, K. Nagasaka, S. Okajima, Y. Tsunawaki, A. Nagashima, A twin optically-pumped far-infrared CH 3 OH laser for plasma diagnostics International Journal of Infrared and Millimeter Waves. ,vol. 1, pp. 57- 76 ,(1980) , 10.1007/BF01007066