摘要: The intraocular pressure (IOP) is an indicator for the diagnosis and treatment of glaucoma. traditional measurement IOP cannot provide a continuous within 24 hours with compromised functionality in glaucoma diagnosis. This paper presents wireless, passive non-invasive contact lens sensor (CLS) monitoring based on flexible substrate. curvature-sensitive includes inductor two capacitors to form C-L-C resonant circuit, which were embedded parylene C using PDMS as intermediate medium layer. circuit used wireless sensing simplifying bonding step. Compared C, has much lower Young Modulus, enabling quantification curvatures, IOP. To fit human eyeballs, size was designed follows: outer diameter: 12 mm, radius curvature: 8.5 mm. characterized vitro obtain frequency response, results show that linearity R>0.97 sensitivity>9.7 kHz/kPa. can be monitor together reading continuously, without impacting daily life users.

参考文章(17)
M. E. Greene, B. G. Gilman, Intraocular Pressure Measurement with Instrumented Contact Lenses Investigative Ophthalmology & Visual Science. ,vol. 13, pp. 299- 302 ,(1974)
Hao Jiang, Di Lan, Hamid Shahnasser, Shuvo Roy, Sensitivity analysis of an implantable LC Based passive sensor biomedical engineering and informatics. ,vol. 4, pp. 1586- 1590 ,(2010) , 10.1109/BMEI.2010.5639234
Guo-Zhen Chen, Ion-Seng Chan, David C.C. Lam, Capacitive contact lens sensor for continuous non-invasive intraocular pressure monitoring Sensors and Actuators A-physical. ,vol. 203, pp. 112- 118 ,(2013) , 10.1016/J.SNA.2013.08.029
Yu-Chieh Huang, Guan-Ting Yeh, Tzu-Sen Yang, Jin-Chern Chiou, A contact lens sensor system with a micro-capacitor for wireless intraocular pressure monitoring ieee sensors. pp. 1- 4 ,(2013) , 10.1109/ICSENS.2013.6688174
R Puers, G Vandevoorde, D De Bruyker, R Puers, G Vandevoorde, Electrodeposited copper inductors for intraocular pressure telemetry Journal of Micromechanics and Microengineering. ,vol. 10, pp. 124- 129 ,(2000) , 10.1088/0960-1317/10/2/305
Matteo Leonardi, Elie M. Pitchon, Arnaud Bertsch, Philippe Renaud, André Mermoud, Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmologica. ,vol. 87, pp. 433- 437 ,(2009) , 10.1111/J.1755-3768.2008.01404.X
E. Hughes, P. Spry, J. Diamond, 24-hour monitoring of intraocular pressure in glaucoma management: a retrospective review. Journal of Glaucoma. ,vol. 12, pp. 232- 236 ,(2003) , 10.1016/J.AJO.2003.10.008
Chien-Kai Tseng, Yu-Chieh Huang, Shang-Wei Tsai, Guan-Ting Yeh, Chung-Hao Chang, Jin-Chern Chiou, Design and fabricate a contact lens sensor with a micro-inductor embedded for intraocular pressure monitoring ieee sensors. pp. 1- 4 ,(2012) , 10.1109/ICSENS.2012.6411234
U Schnakenberg, P Walter, G vom Bögel, C Krüger, H.C Lüdtke-Handjery, H.A Richter, W Specht, P Ruokonen, W Mokwa, Initial investigations on systems for measuring intraocular pressure Sensors and Actuators A-physical. ,vol. 85, pp. 287- 291 ,(2000) , 10.1016/S0924-4247(00)00426-X
Vladimir Laukhin, Irene Sánchez, Ana Moya, Elena Laukhina, Raul Martin, Fernando Ussa, Concepció Rovira, Antón Guimera, Rosa Villa, Jordi Aguiló, José-Carlos Pastor, Jaume Veciana, Non-invasive intraocular pressure monitoring with a contact lens engineered with a nanostructured polymeric sensing film Sensors and Actuators A-physical. ,vol. 170, pp. 36- 43 ,(2011) , 10.1016/J.SNA.2011.05.021