Pure hydroxyapatite phantoms for the calibration of in vivo X-ray fluorescence systems of bone lead and strontium quantification.

作者: Eric Da Silva , Brian Kirkham , Darrick V. Heyd , Ana Pejović-Milić

DOI: 10.1021/AC401877D

关键词:

摘要: Plaster of Paris [poP, CaSO4·(1)/(2) H2O] is the standard phantom material used for calibration in vivo X-ray fluorescence (IVXRF)-based systems bone metal quantification (i.e strontium and lead). Calibration IVXRF employs use a coherent normalization procedure which requires application correction factor (CCF) to data, calculated as ratio relativistic form factors mineral. Various issues have been raised suitability poP include its chemical purity difference from mineral (a calcium phosphate). This work describes preparation chemically pure hydroxyapatite material, known composition stoichiometry, proposed purpose calibrating lead replacement poP. The issue with contamination by t...

参考文章(42)
L. Wielopolski, D. Vartsky, S. Yasumura, S.H. Cohn, Application of XRF to measure strontium in human bone in vivo Advances in x-ray analysis. ,vol. 26, pp. 415- 421 ,(1982) , 10.1154/S0376030800012787
D R Chettle, M C Scott, L J Somervaille, Improvements in the precision of in vivo bone lead measurements. Physics in Medicine and Biology. ,vol. 34, pp. 1295- 1300 ,(1989) , 10.1088/0031-9155/34/9/014
L. Wielopolski, D. N. Slatkin, D. Vartsky, K. J. Ellis, S. H. Cohn, Feasibility Study for the in Vivo Measurement of Lead in Bone Using L-X-Ray Fluorescence IEEE Transactions on Nuclear Science. ,vol. 28, pp. 114- 116 ,(1981) , 10.1109/TNS.1981.4331150
L H Nie, S Sanchez, K Newton, L Grodzins, R O Cleveland, M G Weisskopf, In vivo quantification of lead in bone with a portable x-ray fluorescence system--methodology and feasibility. Physics in Medicine and Biology. ,vol. 56, ,(2011) , 10.1088/0031-9155/56/3/N01
M. Zamburlini, A. Pejović-Milić, D. R. Chettle, Evaluation of geometries appropriate for 125I in vivo bone strontium X-ray fluorescence measurement Journal of Radioanalytical and Nuclear Chemistry. ,vol. 269, pp. 625- 629 ,(2006) , 10.1007/S10967-006-0275-6
Joanne M. O’Meara, Jimmy Börjesson, David R. Chettle, Sören Mattsson, Normalisation with coherent scatter signal: improvements in the calibration procedure of the 57Co-based in vivo XRF bone-Pb measurement. Applied Radiation and Isotopes. ,vol. 54, pp. 319- 325 ,(2001) , 10.1016/S0969-8043(99)00279-1
Dominic Hare, Christine Austin, Philip Doble, Manish Arora, Elemental bio-imaging of trace elements in teeth using laser ablation-inductively coupled plasma-mass spectrometry Journal of Dentistry. ,vol. 39, pp. 397- 403 ,(2011) , 10.1016/J.JDENT.2011.03.004
T. C. Rains, H. E. Zittel, Marion Ferguson, Flame Spectrophotometric Determination of Micro Concentrations of Strontium in Calcareous Materials. Analytical Chemistry. ,vol. 34, pp. 778- 781 ,(1962) , 10.1021/AC60187A019
R E Snyder, D C Secord, The in situ measurement of strontium content in bone using X-ray fluorescence analysis Physics in Medicine and Biology. ,vol. 27, pp. 515- 529 ,(1982) , 10.1088/0031-9155/27/4/003